FilmArray Gastrointestinal (GI) Panel for use with the FilmArray 2.0
Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System. 510(k) K143005, from Biofire Diagnostics, LLC (Salt Lake City, UT). Product code PCH, class 2.
As of Nov 21, 2026, K143005 is an FDA 510(k) held by Biofire Diagnostics, LLC, substantially equivalent on Feb 19, 2015 after 122 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 122 days
- Recalls: 7
- Cited as a predicate by: 4 devices
The FilmArray Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems. The FilmArray GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs…
- Filmarray Gastrointestinal (Gi) PanelK140407 · Biofire Diagnostics · May 2, 2014
- Panther Fusion GI Expanded Bacterial AssayK251993 · Hologic, Inc. · Sep 25, 2025
- Panther Fusion GI Bacterial AssayK251868 · Hologic · Sep 25, 2025
- BD MAX Enteric Viral Panel, BD MAX InstrumentK181427 · Becton, Dickinson and Company · Nov 29, 2018
- FilmArray Gastrointestinal Panel (GI) for use with FilmArray TorchK160459 · Biofire Diagnostics, LLC · Apr 1, 2016
| May 16, 2024 | Z-2514-2024Due to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium. | Open |
| Jan 26, 2024 | Z-1140-2024Due to potential signals of increased false positive Norovirus results when using the gastrointestinal (GI) panel. | Open |
| Nov 27, 2023 | Z-0703-2024A manufacturing event may contribute to elevated rates of false negative results while using gastrointestinal panel and respiratory panel, which could lead to inappropriate antibiotic treatment, no treatment, delay of… | Open |
| Apr 25, 2023 | Z-2588-2023The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel | Open |
| Mar 8, 2023 | Z-1392-2023Due to manufacturing issue, Gastrointestinal (GI) Panel may result in false negative results. | Open |
| Feb 1, 2023 | Z-1253-2023Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping… | Open |
| Aug 8, 2019 | Z-1451-2020Complaints have been received concerning elevated rates of false negative results while using the Gastrointestinal (GI) Panel. False negative results may lead to incorrect treatment being provided which could… | Terminated |
- BIOFIRE FILMARRAY Gastrointestinal 1.1 (GI1.1) Panel; BIOFIRE FILMARRAY Gastrointestinal 1.1 (GI1.1) Panel MidK262172 · Biofire Diagnostics, LLC · Nov 6, 2026
- LIAISON PLEX Gastrointestinal Flex AssayK253722 · Luminex Corporation · May 19, 2026
- QIAstat-Dx Gastrointestinal Panel 2; QIAstat-Dx GI Panel 2 Mini B&V; QIAstat-Dx GI Panel 2 Mini BK254032 · QIAGEN GmbH · Mar 9, 2026
- Xpert GI PanelK251721 · Cepheid · Jan 16, 2026
- QIAstat-Dx Gastrointestinal Panel 2; QIAstat-Dx GI Panel 2 Mini B&V; QIAstat-Dx GI Panel 2 Mini BK252329 · QIAGEN GmbH · Oct 22, 2025
- Panther Fusion GI Bacterial AssayK251868 · Hologic · Sep 25, 2025
- Panther Fusion GI Expanded Bacterial AssayK251993 · Hologic, Inc. · Sep 25, 2025
- QIAstat-Dx GI Panel 2 Mini BK250324 · QIAGEN GmbH · Feb 28, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 20, 2014
- Review panel
- Microbiology
- Regulation
- 21 CFR 866.3990
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files