Filmarray Gastrointestinal (Gi) Panel
Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System. 510(k) K140407, from Biofire Diagnostics (Salt Lake City, UT). Product code PCH, class 2.
As of Nov 20, 2026, K140407 is an FDA 510(k) held by Biofire Diagnostics, substantially equivalent on May 2, 2014 after 73 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 73 days
- Recalls: 8
- Cited as a predicate by: 8 devices
- K121454K121454
- Panther Fusion GI Expanded Bacterial AssayK251993 · Hologic, Inc. · Sep 25, 2025
- Panther Fusion GI Bacterial AssayK251868 · Hologic · Sep 25, 2025
- Xpert GI PanelK251721 · Cepheid · Jan 16, 2026
- QIAstat-Dx Gastrointestinal Panel 2K220062 · QIAGEN GmbH · May 31, 2024
- EntericBio Dx AssayK182703 · Serosep , Ltd. · Jun 19, 2019
- Great Basin Shiga Toxin Direct TestK152955 · Great Basin Scientific, Inc. · Mar 22, 2016
- BD MAX Enteric Parasite Panel, BD MAX InstrumentK143648 · Becton, Dickinson and Company · Aug 25, 2015
- FilmArray Gastrointestinal (GI) Panel for use with the FilmArray 2.0K143005 · Biofire Diagnostics, LLC · Feb 19, 2015
| May 16, 2024 | Z-2514-2024Due to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium. | Open |
| Jan 26, 2024 | Z-1140-2024Due to potential signals of increased false positive Norovirus results when using the gastrointestinal (GI) panel. | Open |
| Nov 27, 2023 | Z-0703-2024A manufacturing event may contribute to elevated rates of false negative results while using gastrointestinal panel and respiratory panel, which could lead to inappropriate antibiotic treatment, no treatment, delay of… | Open |
| Apr 25, 2023 | Z-2588-2023The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel | Open |
| Mar 8, 2023 | Z-1392-2023Due to manufacturing issue, Gastrointestinal (GI) Panel may result in false negative results. | Open |
| Feb 1, 2023 | Z-1253-2023Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping… | Open |
| Aug 8, 2019 | Z-1451-2020Complaints have been received concerning elevated rates of false negative results while using the Gastrointestinal (GI) Panel. False negative results may lead to incorrect treatment being provided which could… | Terminated |
| Mar 6, 2019 | Z-1242-2019Elevated rates of false positive results for Campylobacter and Cryptosporidium have been identified. | Terminated |
- BIOFIRE FILMARRAY Gastrointestinal 1.1 (GI1.1) Panel; BIOFIRE FILMARRAY Gastrointestinal 1.1 (GI1.1) Panel MidK262172 · Biofire Diagnostics, LLC · Nov 6, 2026
- LIAISON PLEX Gastrointestinal Flex AssayK253722 · Luminex Corporation · May 19, 2026
- QIAstat-Dx Gastrointestinal Panel 2; QIAstat-Dx GI Panel 2 Mini B&V; QIAstat-Dx GI Panel 2 Mini BK254032 · QIAGEN GmbH · Mar 9, 2026
- Xpert GI PanelK251721 · Cepheid · Jan 16, 2026
- QIAstat-Dx Gastrointestinal Panel 2; QIAstat-Dx GI Panel 2 Mini B&V; QIAstat-Dx GI Panel 2 Mini BK252329 · QIAGEN GmbH · Oct 22, 2025
- Panther Fusion GI Bacterial AssayK251868 · Hologic · Sep 25, 2025
- Panther Fusion GI Expanded Bacterial AssayK251993 · Hologic, Inc. · Sep 25, 2025
- QIAstat-Dx GI Panel 2 Mini BK250324 · QIAGEN GmbH · Feb 28, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 18, 2014
- Review panel
- Microbiology
- Regulation
- 21 CFR 866.3990
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files