Arca
Skip to content
Tools
Tools

Parietex Monofilament Polyester Mesh (new name: Parietex Lightweight Mesh), Parietex Composite Mono PM Mesh (new name: Parietex Composite Parastomal Mesh), Parietex Composite Ventral Patch, Symbotex Composite Mesh

Mesh, Surgical, Polymeric. 510(k) K142908, from Sofradim Production (Trévoux, FR). Product code FTL, class 2.

As of Nov 19, 2026, K142908 is an FDA 510(k) held by Sofradim Production, substantially equivalent on Nov 14, 2014 after 39 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 39 days
  • Recalls: 1
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

ParietexTM lightweight mesh is indicated for inguinal and ventral hernia repair. ☒

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K142908, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Oct 5, 2018Z-0461-2019The firm is issuing this voluntary removal following receipt of reports of parastomal mesh failure identified several years following parastomal hernia repair using the modified Sugarbaker repair technique. In these…Terminated

Other FTL clearances

The newest 510(k)s for the same product code; their median review was 85 days; see every FTL device

Record

Type
Special
Received
Oct 6, 2014
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.3300

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.