Optima XR200amx, Optima XR220amx

System, X-Ray, Mobile. 510(k) K142383, from Ge Medical Systems, LLC (Waukesha, WI). Product code IZL, class 2.

As of Nov 20, 2026, K142383 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Sep 25, 2014 after 30 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 30 days
  • Recalls: 8
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Optima XR200amx and Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K142383, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 5, 2026Z-2727-2026, Z-2726-2026GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly.…Open
Nov 25, 2024Z-0811-2025, Z-0812-2025GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and…Open
Dec 23, 2015Z-0770-2016, Z-0768-2016Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result…Terminated
Dec 19, 2014Z-1406-2015, Z-1407-2015An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec.Terminated

Other IZL clearances

The newest 510(k)s for the same product code; their median review was 166 days; see every IZL device

Record

Type
Special
Received
Aug 26, 2014
Review panel
Radiology
Regulation
21 CFR 892.1720

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files