BRIVOXR285AMX, Optima XR200AMX, OPTIMAXR220AMX

System, X-Ray, Mobile. 510(k) K103476, from Ge Medical Systems, LLC (Waukesha, WI). Product code IZL, class 2.

As of Nov 17, 2026, K103476 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Jun 16, 2011 after 204 days, with 8 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jun 16, 2011
  • Review time: 204 days
  • Recalls: 8
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx, Optima XR200amx, Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K103476, newest first; each opens FDA's recall record
Jun 5, 2026Z-2727-2026GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly.…Open, Classified
Jun 5, 2026Z-2726-2026GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly.…Open, Classified
Nov 25, 2024Z-0811-2025GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and…Open, Classified
Nov 25, 2024Z-0812-2025GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and…Open, Classified
Dec 23, 2015Z-0769-2016Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result…Terminated
Dec 19, 2014Z-1405-2015An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec.Terminated
Nov 7, 2013Z-0725-2014There exists a software issue associated with patient selection from the worklist on the Optima XR200 with Digital Upgrade, and Optima XR220amx Mobile X-ray Systems. Intermittently when a large number of patient exams…Terminated
Aug 7, 2012Z-1942-2013GE Healthcare conducted a recall on various mobile x-ray units (Optima XR220amx, Optima XR200amx, and Brivo XR285amx systems) because of an Erratic Drive Performance.Terminated

Other IZL Clearances

The newest 510(k)s for the same product code; their median review was 166 days; see every IZL device

Record

Type
Traditional
Received
Nov 24, 2010
Review panel
Radiology
Regulation
21 CFR 892.1720

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.