Artis Zee/Zeego SW VC21

Interventional Fluoroscopic X-Ray System. 510(k) K141574, from Siemens Medical Solutions USA, Inc. (Mavern, PA). Product code OWB, class 2.

As of Nov 19, 2026, K141574 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Sep 5, 2014 after 85 days, with 12 recalls tied to it in FDA's recall database.

  • Review time: 85 days
  • Recalls: 12
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Artis zee / zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K141574, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 25, 2018Z-2526-2018Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedureTerminated
May 1, 2018Z-2309-2018, Z-2313-2018, Z-2310-2018, Z-2312-2018, Z-2311-2018, Z-2314-2018, Z-2308-2018After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure.…Terminated
Sep 28, 2017Z-0710-2018The possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be…Terminated
Dec 8, 2016Z-0971-2017Siemens initiated a corrective action to address two possible, mutually independent causes of a system defect related to the following: - In Artis Systems with A100 Plus or A100G generators, an attempt to resume…Terminated
Nov 1, 2016Z-0602-2017The possibility exists that the semi-automatic focus switch may not work as intended when pressing the footswitch multiple times.Terminated
May 22, 2015Z-1846-2015There exists a possible position sensor fault in the swivel base axis not being detected by the system software. When moving, the system could potentially exceed its usual speed, resulting in a collision. No injuries…Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Special
Received
Jun 12, 2014
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files