Arca
Skip to content
Tools
Tools

Flow-I Anesthesia System C20, C30 & C40

Gas-Machine, Anesthesia. 510(k) K133958, from Maquet Critical Care AB (Wayne, NJ). Product code BSZ, class 2.

As of Nov 19, 2026, K133958 is an FDA 510(k) held by Maquet Critical Care AB, substantially equivalent on Mar 20, 2014 after 86 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 86 days
  • Recalls: 8
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The indication for FLOW-i Anesthesia System is administering inhalation Anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe. The system is intended for use on neonatal to adult patient populations.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K133958, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 11, 2023Z-0989-2024, Z-0988-2024, Z-0987-2024, Z-0991-2024, Z-0990-2024Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous…Open
Dec 10, 2015Z-0804-2016Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the sub-systems and, in a few…Terminated
Apr 2, 2015Z-1706-2015In some Flow-i Anesthesia Systems, patient cassette can get dislodged which could result in gas leakage within the system internal circuit potentially leading to a stop in ventilation.Terminated
Mar 2, 2015Z-2083-2015Maquet has identified that in some FLOW-i Anesthesia System Auxiliary O2 and Suction Modules, the suctioning function cannot be activated or controlled by the suction unit on/off switch leading to a decrease or loss of…Terminated

Other BSZ clearances

The newest 510(k)s for the same product code; their median review was 265 days; see every BSZ device

Record

Type
Abbreviated
Received
Dec 24, 2013
Review panel
Anesthesiology
Regulation
21 CFR 868.5160

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files