Artis One

Interventional Fluoroscopic X-Ray System. 510(k) K133580, from Siemens Medical Solutions USA, Inc. (Mavern, PA). Product code OWB, class 2.

As of Nov 18, 2026, K133580 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Apr 28, 2014 after 158 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 158 days
  • Recalls: 7
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis one include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whold body radiographic/fluoroscopic procedures as well as…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K133580, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 2, 2025Z-0922-2025A potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue…Open
Mar 21, 2023Z-1469-2023Examination room monitor connected with the hybrid cable may have a display issue and no image function is availableOpen
Sep 21, 2020Z-0298-2021Potential malfunction in which the examination room monitor may become intermittently dysfunctional (e.g., no display/flicking/wrong resolution).Terminated
Jan 12, 2018Z-1148-2018Certain patient table systems may have a missing or incorrectly installed snap ring that could cause the bolt to slide out of its base and the tabletop carriage to slide to the floor.Terminated
Aug 22, 2017Z-0614-2018The possibility exists for the monitor display of Artis one systems with a DP Transceiver below revision 03 to experience blinking and/or a blue/black screenTerminated
Mar 11, 2016Z-1282-2016After importing segmentation results of the left atrium created on the Artis One system, the possibility exists for the results to be mirrored to an electro-anatomical 3D mapping system, CARTO from Biosense Webster Inc.Terminated
Jan 16, 2015Z-1119-2015The possibility exists that the monitor may fail and requires a power circle (shutdown and then power on) to resume operation. The problem is not systematic; but sporadic on single units.Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Nov 21, 2013
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files