Alaris System With Guardrails Suite Mx

Pump, Infusion. 510(k) K133532, from Carefusion 303, Inc. (San Diego, CA). Product code FRN, class 2.

As of Nov 18, 2026, K133532 is an FDA 510(k) held by Carefusion 303, Inc., substantially equivalent on Aug 21, 2014 after 276 days, with 25 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Aug 21, 2014
  • Review time: 276 days
  • Recalls: 25+
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Alaris System with Guardrails Suite MX is intended for use in professional healthcare facilities that utilize infusion devices for the delivery of fluids, medications, blood and blood products. The Alaris System with Guardrails Suite MX is intended to provide trained healthcare caregivers a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps…

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K133532, newest first; the 25 newest; each opens FDA's recall record
Feb 11, 2026Z-1645-2026Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care UnitsOpen, Classified
Sep 15, 2023Z-0186-2024Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.Open, Classified
Sep 15, 2023Z-0185-2024Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.Open, Classified
Sep 15, 2023Z-0184-2024Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.Open, Classified
Mar 26, 2021Z-1544-2021Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel…Open, Classified
Mar 25, 2021Z-1651-2021Bezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts…Open, Classified
Mar 24, 2021Z-1375-2021Aftermarket front bezel components were installed during service/repair, not by the original manufacturer, within infusion pump modules. Over time, the posts on the front bezel component may crack or separate.…Open, Classified
Mar 3, 2021Z-1309-2021Pump Module keypad lifting, and Fluid ingress could result in 1) Unresponsive keys: module continues infusion, PC unit will not alarm and must be used for programming changes, may necessitate a different Pump Module.…Open, Classified
Feb 26, 2021Z-1543-2021Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel…Terminated
Feb 25, 2021Z-1332-2021Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel…Terminated
Feb 25, 2021Z-2204-2021Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel…Terminated
Feb 25, 2021Z-1331-2021Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel…Terminated
Aug 4, 2020Z-2882-2020The Alaris PC unit can display incorrect syringe type and/or syringe sizes. This could result in delays in infusion, under-infusion, or over-infusion.Open, Classified
Aug 4, 2020Z-2939-2020Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or…Open, Classified
Aug 4, 2020Z-2909-2020The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate…Open, Classified
Aug 4, 2020Z-2935-2020Infusion pump component defect may result in interruption of patient monitoringOpen, Classified
Aug 4, 2020Z-2910-2020The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate…Open, Classified
Jun 30, 2020Z-2741-2020If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged,…Terminated
Jun 30, 2020Z-2736-2020Broken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dropped, or if improper cleaners are used…Terminated
Jun 30, 2020Z-2822-2020LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim.Open, Classified
Jun 30, 2020Z-2824-2020LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim.Open, Classified
Jun 30, 2020Z-2716-2020Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL…Terminated
Jun 30, 2020Z-2823-2020LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim.Open, Classified
Jun 30, 2020Z-2720-2020Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL…Terminated
Jun 30, 2020Z-2740-2020If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged,…Terminated

Other FRN Clearances

The newest 510(k)s for the same product code; their median review was 242 days; see every FRN device

Record

Type
Traditional
Received
Nov 18, 2013
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files