Arca
Skip to content
Tools
Tools

Mega 8FR. 50CC and Sensation Plus 8FR. 50CC Intra-Aortic Balloon Catheters and Accessories

System, Balloon, Intra-Aortic And Control. 510(k) K133074, from Datascope Corp. (Fairfield, NJ). Product code DSP, class 2.

As of Nov 28, 2026, K133074 is an FDA 510(k) held by Datascope Corp., substantially equivalent on Dec 12, 2013 after 73 days, with 9 recalls tied to it in FDA's recall database.

  • Review time: 73 days
  • Recalls: 9

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Refractory Unstable Angina. - Impending Infarction. • Acute Myocardial Infarction. • Refractory Ventricular Failure. • Complications of Acute MI (le. Acute MR or VSD or papillary muscle rupture) • Cardiogenic Shock. - Support for diagnostic, percutaneous revascularization and interventional procedures. • Ischemia related intractable ventricular arrhythmias. - Septic Shock.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K133074, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 17, 2024Z-0311-2025There is discrepant labeling on the inner and outer packaging of the device.Open
Jun 23, 2023Z-2220-2023, Z-2222-2023During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta,…Open
Aug 5, 2022Z-1687-2022, Z-1685-2022, Z-1686-2022Certain lots containing undersized dilator.Open
Jul 27, 2020Z-2837-2020Potential Endotoxin ContaminationOpen
Feb 21, 2018Z-1385-2018The U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section.Terminated
Apr 12, 2017Z-2164-2017Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing…Terminated

Other DSP clearances

The newest 510(k)s for the same product code; see every DSP device

Record

Type
Special
Received
Sep 30, 2013
Review panel
Cardiovascular
Regulation
21 CFR 870.3535

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files