Screw, Fixation, Bone. 510(k) K131920, from In2bones Sas (Potomac, MD). Product code HWC, class 2.
As of Nov 18, 2026, K131920 is an FDA 510(k) held by In2bones Sas, substantially equivalent on May 1, 2014 after 309 days, with 3 recalls tied to it in FDA's recall database.
- Review time: 309 days
- Recalls: 3
- Cited as a predicate by: 4 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe I.B.S.TM osteosynthesis screws are intended for: - The fixation of arthrodeses, osteotomies or fractures of long or short bones of the upper and lower limbs and lower limbs - Osteosyntheses requiring a mono or bicortical compression The size of the chosen screw should be adapted to the specific indications.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Stabilization ScrewK050346 · New Deal, S.A. · Mar 17, 2005
- Sbi AutofixK052576 · Small Bone Innovations, LLC · Nov 7, 2005
- Synthes 7.0/7.3 MM Cannulated ScrewsK962011 · Synthes (Usa) · Aug 5, 1996
- Qwix Positioning ScrewK071639 · Newdeal Sas · Jul 9, 2007
- Acutrak Plus Fixation SystemK944330 · Acu Med, Inc. · May 5, 1995
Cited as a predicate
4 later 510(k) summaries name this device as a predicate- F.A.S.T ScrewK254280 · Steps Ortho · Sep 21, 2026
- I.B.S. osteosynthesis screwsK170594 · In2bones Sas · May 10, 2017
- In2Bones Fracture and Correction SystemK170518 · In2bones USA, LLC · Jul 14, 2017
- I.B.S. 2.0 Osteosynthesis screwK160174 · In2bones Sas · Oct 6, 2016
Recalls
Recalls FDA ties to K131920, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Jan 26, 2024 | Z-1320-2024, Z-1319-2024This Field Action is being conducted following the identification of a batch mix-up between I.B.S Compression screws diameter 6.5mm length 75mm and I.B.S Compression screws diameter 6.5mm length 80mm. The screws of the… | Open |
| Nov 18, 2015 | Z-0397-2016One batch of osteosynthesis screws were manufactured with the incorrect Torx imprint on the screw head. | Terminated |
Other HWC clearances
The newest 510(k)s for the same product code; their median review was 157 days; see every HWC device- Ntegrate Threaded Nail SystemK261961 · Nvision Biomedical Technologies · Oct 14, 2026
- OsteoCentric Technologies Cannulated and Soild Fasteners with NutsK263116 · OsteoCentric Technologies · Oct 2, 2026
- Arthrex Beaming SystemK262975 · Arthrex, Inc. · Sep 25, 2026
- DynaNail Helix® Fixation SystemK260995 · MedShape, Inc. · Sep 23, 2026
- F.A.S.T ScrewK254280 · Steps Ortho · Sep 21, 2026
- ActivaScrew™ Cannulated COMK261057 · Bioretec, Ltd. · Sep 4, 2026
- OsteoCoil™ GT Active Compression SystemK260106 · Flower Orthopedics Corporation DBA Conventus Flower Orthoped · Aug 24, 2026
- Large Compression ScrewsK262344 · TriMed, Inc. · Aug 12, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 26, 2013
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3040
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files