Stabilization Screw
Screw, Fixation, Bone. 510(k) K050346, from New Deal, S.A. (Planisboro, NJ). Product code HWC, class 2.
As of Nov 17, 2026, K050346 is an FDA 510(k) held by New Deal, S.A., substantially equivalent on Mar 17, 2005 after 34 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Mar 17, 2005
- Review time: 34 days
- Recalls: 4
- Cited as a predicate by: 5 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe STABILIZATION SCREW is indicated for fixation of bone fractures or for bone reconstruction. Examples include: - Mono or Bi-cortical osteotomies in the foot or hand (including Hallux Valgus treatment) - Fractures management in the foot or hand - Fixation of bone fragments in long bones or small bones fractures - Arthrodesis in hand, foot or ankle surgery The size of the chosen screw should be…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Modification To The Bold ScrewK011262 · New Deal, S.A. · Jun 15, 2001
- Scarf Thread-Head (Tm) Head ScrewK971070 · Landos, Inc. · Jul 9, 1997
- Modification To The I.C.O.S ScrewK011821 · New Deal, S.A. · Sep 6, 2001
- Synthes Sterile 4.5 Mm Cannulated ScrewsK963172 · Synthes (Usa) · Nov 8, 1996
Cited as a Predicate
5 later 510(k) summaries name this device as a predicate- Large Screws RangeK160617 · Newclip Technics · Jun 3, 2016
- Stabilization ScrewK131920 · In2bones Sas · May 1, 2014
- Ars Screw; Slant ScrewK130954 · Advanced Interventional Technology, LLC · May 31, 2013
- Merete Compression ScrewsK091798 · Merete Medical GmbH · Sep 2, 2009
- Qwix Positioning ScrewK071639 · Newdeal Sas · Jul 9, 2007
Recalls
Recalls FDA ties to K050346, newest first; each opens FDA's recall record| Feb 8, 2010 | Z-1254-2010The screws were etched incorrectly as 32mm instead of the correct 34mm length. | Terminated |
| Feb 8, 2010 | Z-1253-2010The screws were etched incorrectly as 34mm instead of the correct 32mm. | Terminated |
| Feb 27, 2008 | Z-1445-2008Certain QWIX Fixation Screws (Reference numbers 111426S and 111428S) have been etched and labelled with an incorrect length. | Terminated |
| Feb 27, 2008 | Z-1446-2008Certain QWIX Fixation Screws (Part numbers 111426s and 111428s) have been etched and labelled with an incorrect length. | Terminated |
Other HWC Clearances
The newest 510(k)s for the same product code; their median review was 157 days; see every HWC device- Ntegrate Threaded Nail SystemK261961 · Nvision Biomedical Technologies · Oct 14, 2026
- OsteoCentric Technologies Cannulated and Soild Fasteners with NutsK263116 · OsteoCentric Technologies · Oct 2, 2026
- Arthrex Beaming SystemK262975 · Arthrex, Inc. · Sep 25, 2026
- DynaNail Helix® Fixation SystemK260995 · MedShape, Inc. · Sep 23, 2026
- F.A.S.T ScrewK254280 · Steps Ortho · Sep 21, 2026
- ActivaScrew Cannulated COMK261057 · Bioretec, Ltd. · Sep 4, 2026
- OsteoCoil GT Active Compression SystemK260106 · Flower Orthopedics Corporation DBA Conventus Flower Orthoped · Aug 24, 2026
- Large Compression ScrewsK262344 · TriMed, Inc. · Aug 12, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 11, 2005
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3040
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files