Ventilator, Continuous, Facility Use. 510(k) K131774, from Hamilton Medical AG (Bonaduz, CH). Product code CBK, class 2.

As of Nov 17, 2026, K131774 is an FDA 510(k) held by Hamilton Medical AG, substantially equivalent on Oct 9, 2013 after 114 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Oct 9, 2013
  • Review time: 114 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K131774, newest first; each opens FDA's recall record
Mar 25, 2019Z-1874-2019New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air…Terminated
Mar 9, 2016Z-1624-2016After performing the suctioning maneuver, including disconnecting the patient, suctioning , and reconnecting the patient, the preset pattern of ventilation many not continue as expected.Terminated
Mar 24, 2015Z-2057-2015Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device.Terminated
Apr 22, 2014Z-0151-2016Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the operator; this situation can occur regardless of the selected patient group…Terminated

Other CBK Clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Traditional
Received
Jun 17, 2013
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files