HAMILTON-G5
Ventilator, Continuous, Facility Use. 510(k) K131774, from Hamilton Medical AG (Bonaduz, CH). Product code CBK, class 2.
As of Nov 17, 2026, K131774 is an FDA 510(k) held by Hamilton Medical AG, substantially equivalent on Oct 9, 2013 after 114 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Oct 9, 2013
- Review time: 114 days
- Recalls: 4
- Cited as a predicate by: 3 devices
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- HAMILTON-G5K103803 · Hamilton Medical AG · Nov 23, 2011
- Servo-I Ventilator System Model 64 87 800 E4073, Heliox Option, Model 6675585K073179 · Maquet Critical Care AB · Jun 9, 2008
- Masimo Rainbow Set Radical 7R Pulse Co-OximeterK100428 · Masimo Corporation · Jul 9, 2010
- Nihon Kohden TL-101T, TL-120T And TL-121T SPO2 Probes And AccessoriesK974292 · Nihon Kohden America, Inc. · Jun 8, 1998
- Nihon Kohden BSM-2300A Series Bedside Monitor And Accessories, Model BSM-2300A SeriesK011918 · Nihon Kohden America, Inc. · Sep 12, 2001
- Bedside Monitor, Model OPV-1500K032749 · Nihon Kohden America, Inc. · Sep 16, 2003
- Transmitter, ZS-940PA SeriesK043517 · Nihon Kohden America, Inc. · Feb 3, 2005
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- Hamilton-G5K180295 · Hamiton Medical AG · Jul 30, 2019
- HAMILTON-C3K161450 · Hamilton Medical AG · Apr 27, 2017
- HAMILTON-T1,HAMILTON-C1K140939 · Hamilton Medical AG · Sep 2, 2015
Recalls
Recalls FDA ties to K131774, newest first; each opens FDA's recall record| Mar 25, 2019 | Z-1874-2019New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air… | Terminated |
| Mar 9, 2016 | Z-1624-2016After performing the suctioning maneuver, including disconnecting the patient, suctioning , and reconnecting the patient, the preset pattern of ventilation many not continue as expected. | Terminated |
| Mar 24, 2015 | Z-2057-2015Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device. | Terminated |
| Apr 22, 2014 | Z-0151-2016Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the operator; this situation can occur regardless of the selected patient group… | Terminated |
Other CBK Clearances
The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device- Evita (V800); Evita (V600)K253647 · Draegerwerk AG & CO Kgaa · May 7, 2026
- Inbentus VersatileK252663 · Iconic Solutions by Murcia, SL · Apr 14, 2026
- VPAP Pediatric Face MaskK251888 · Hsiner Co., Ltd. · Mar 12, 2026
- Vivo 45 LSK232365 · Oconnell Regulatory Consultants, Inc. · Dec 19, 2025
- F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits Size A…K252173 · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
- SV600, SV800 VentilatorK243767 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 28, 2025
- AVEA disposable expiratory filter/water trap (11790)K251133 · Telesair, Inc. · May 8, 2025
- AF531 Oro-Nasal SE Face MaskK243394 · Respironics, Inc. · Dec 30, 2024
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 17, 2013
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files