Centrimag Primary Console (2ND Generation) and 1 more
Control, Pump Speed, Cardiopulmonary Bypass. 510(k) K131179, from Thoratec Corp (Burlington, MA). Product code DWA, class 2.
As of Nov 19, 2026, K131179 is an FDA 510(k) held by Thoratec Corp, substantially equivalent on Jul 18, 2013 after 84 days, with 1 recall tied to it in FDA's recall database.
- Review time: 84 days
- Recalls: 1
- Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100K020271 · Levitronix, LLC · Mar 25, 2003
- Levitronix 2ND Generation Centrimag Primary Console, Levitronix MonitorK102129 · Levitronix, LLC · Aug 27, 2010
| Started | Recall | Status |
|---|---|---|
| Sep 5, 2018 | Z-0103-2019Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection. | Terminated |
- Inspira ART100K232788 · Inspira Technologies Oxy B.H.N. , Ltd. · May 24, 2024
- Anivia SG1000 Pump ConsoleK230698 · Apmtd, Inc. · Apr 18, 2023
- Anivia SG1000 Pump ConsoleK221491 · Apmtd, Inc. · Feb 3, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Apr 25, 2013
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.4380
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files