Arca
Skip to content
Tools
Tools

Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100

Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type. 510(k) K020271, from Levitronix, LLC (Waltham, MA). Product code KFM, class 2.

As of Nov 19, 2026, K020271 is an FDA 510(k) held by Levitronix, LLC, substantially equivalent on Mar 25, 2003 after 421 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 421 days
  • Recalls: 3
  • Cited as a predicate by: 10 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

CentriMag® Extracorporeal Blood Pumping System The Levitronix CentriMag® Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours).

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

10 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K020271, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 22, 2019Z-0221-2020Reports have been received of the CentriMag Systems experiencing motor and pump issues resulting from electromagnetic interference (EMI).Terminated
Sep 5, 2018Z-0103-2019Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection.Terminated
Mar 17, 2008Z-1901-2008Stops pumping: Interruption of CentriMag System support may occur when using a Valleylab Force FX-C Electrocautery UnitTerminated

Other KFM clearances

The newest 510(k)s for the same product code; their median review was 422 days; see every KFM device

Record

Type
Abbreviated
Received
Jan 28, 2002
Review panel
Cardiovascular
Regulation
21 CFR 870.4360

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files