Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100
Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type. 510(k) K020271, from Levitronix, LLC (Waltham, MA). Product code KFM, class 2.
As of Nov 19, 2026, K020271 is an FDA 510(k) held by Levitronix, LLC, substantially equivalent on Mar 25, 2003 after 421 days, with 3 recalls tied to it in FDA's recall database.
- Review time: 421 days
- Recalls: 3
- Cited as a predicate by: 10 devices
CentriMag® Extracorporeal Blood Pumping System The Levitronix CentriMag® Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours).
- BPX80 Bio-Pump (BPX80)K973011 · Medtronic Bio-Medicus, Inc. · Feb 25, 1998
- Bio-Medicus Bio-Console Models 540, 540A,B,C,D,E, & 550K941921 · Medtronic Bio-Medicus, Inc. · Jul 20, 1994
- CentriMag Acute Circulatory Support SystemK191557 · Thoratec Corporation (Now Part of Abbott) · Jul 10, 2019
- Centrimag Primary Console (2ND Generation); Mag MonitorK131179 · Thoratec Corp · Jul 18, 2013
- Tandemheart PumpK110493 · Cardiacassist, Inc. · Sep 20, 2011
- Cardiohelp SystemK102726 · Maquet Cardiopulmonary, AG · Apr 4, 2011
- Levitronix 2ND Generation Centrimag Primary Console, Levitronix MonitorK102129 · Levitronix, LLC · Aug 27, 2010
- Levitronix Pedimag Blood Pump, Model 201-90052K090051 · Levitronix, LLC · Oct 8, 2009
- Levitronix Centrimag Primary Console, Thoratec Centrimag Primary ConsoleK081221 · Levitronix, LLC · Jun 19, 2008
- Orqis Extracorporeal Blood Pumping System, Model BPS-100K073631 · Orqis Medical Corporation · Jun 27, 2008
- Levitronix Centrimag Primary ConsoleK053630 · Levitronix, LLC · Jan 26, 2006
- Levitronix Centrimag Back-Up ConsoleK051209 · Levitronix, LLC · Sep 8, 2005
| Aug 22, 2019 | Z-0221-2020Reports have been received of the CentriMag Systems experiencing motor and pump issues resulting from electromagnetic interference (EMI). | Terminated |
| Sep 5, 2018 | Z-0103-2019Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection. | Terminated |
| Mar 17, 2008 | Z-1901-2008Stops pumping: Interruption of CentriMag System support may occur when using a Valleylab Force FX-C Electrocautery Unit | Terminated |
- Quantum Perfusion Centrifugal Blood Pump CP20NG with Integrated Sensors (CP20V-NG)K251762 · Spectrum Medical S.R.L. · Aug 4, 2025
- Quantum Perfusion Centrifugal Blood Pump CP22NG with Integrated Sensors (CP22V-NG)K250326 · Spectrum Medical S.R.L. · Jun 24, 2025
- Quantum Perfusion Centrifugal Blood Pump CP37 with Integrated Sensors (CP37V-VT); Quantum Perfusion Centrifugal Blood…K233091 · Qura S.R.L · Oct 26, 2023
- LifeSPARC SystemK232132 · Cardiacassist, Inc. · Aug 3, 2023
- Quantum Perfusion Centrifugal Blood Pump CP22 with Integrated Sensors, Quantum Perfusion Centrifugal Blood Pump CP20…K220842 · Qura S.R.L · May 19, 2023
- Capiox iCP Centrifugal PumpK200091 · Terumo Cardiovascular Systems Corporation · Nov 18, 2021
- Quantum PureFlow Centrifugal Blood Pump CP22 with Integrated SensorK201320 · Qura S.R.L · Nov 3, 2021
- FloPump 57mL Centrifugal PumpK193663 · International Biophysics Corporation · Mar 10, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Abbreviated
- Received
- Jan 28, 2002
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.4360
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files