System, Imaging, Pulsed Doppler, Ultrasonic. 510(k) K123622, from Siemens Medical Solutions USA, Inc. (Moutain View, CA). Product code IYN, class 2.

As of Nov 19, 2026, K123622 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Dec 5, 2012 after 12 days, with 13 recalls tied to it in FDA's recall database.

  • Review time: 12 days
  • Recalls: 13
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Ultrasound imaging or fluid flow analysis of the human body as follows: Clinical Application A B M PWD CWD Color Doppler Power (Amplitude) Doppler Color Velocity Imaging Combined (Specify) Other: Harmonic Imaging Ophthalmic . Fetal Abdominal Intraoperative Abdominal . Intraoperative Neurological P P P P P P P* P Pediatric . Small Organ (specify) . . Neonatal Cephalic Adult Cephalic Cardiac .

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K123622, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 7, 2016Z-0667-2017Firm received reports about "acquisition errors" and the deterioration of material covering the articulating section of the Z6Ms transesophageal transducer.Terminated
Apr 8, 2016Z-1592-2016While imaging with a transesophageal (TEE) transducer (Z6Ms, V5Ms, or V7M), user may lose the ability to control the color region of interest, the pulsed wave or continuous wave Doppler gate or cursor, the M-mode…Terminated
Oct 7, 2015Z-0182-2016Siemens will provide all customers a new version of system software (VA35E or VB10C depending on the hardware configuration). Hardware fixes will be implemented to address situations where the metal component used to…Terminated
Aug 19, 2015Z-2783-2015The ACUSON SC2000 ultrasound system considers uppercase/lowercase differences in the same patient name as unique patient instances when registered on the same ultrasound system. If these differences are not corrected at…Terminated
Jan 12, 2015Z-1056-2015The bolt holding the control panel of the ACUSON SC2000 in a fixed position may fail and the Control panel cannot be locked into position.Terminated
Apr 23, 2014Z-1667-2014Software issues with ACUSON SC2000 ultrasound with software version 3/5 (VA35) - may incorrectly display the color flow data in the wrong position and when Doppler Frequency is adjusted, the system does not restart the…Terminated
Apr 23, 2014Z-1668-2014Potential malfunction due to a software issue for ACUSON SC2000 ultrasound systems using software version 3.5 (VA35) when using the stress echo feature which could cause the need for the study to be repeated.Terminated
Dec 22, 2013Z-0824-2014The user interface assembly on the ACUSON SC2000 may become loose with the potential for the entire module to fall off of the ultrasound system.Terminated
Dec 22, 2013Z-0825-2014The locking mechanism that is intended to hold the control panel in a fixed position can become loose and fail to lock the rotation of the control panel, making it unable to steer or control during transport.Terminated
Jul 10, 2013Z-0052-2014Software issue results in inaccurate 3D image may result in ablation in wrong location.Terminated
Mar 28, 2013Z-1081-2013When using the V5M transducer, the display of the temperature on the ACUSON SC2000 system can be lower than the actual temperature of the V5M transducer. This occurs only if the system hardware contains revision 4 of…Terminated
Jul 11, 2012Z-1584-2015In certain situations, loss of post exercise images can occur in Stress Echo.Terminated
May 11, 2012Z-1581-2015In some cases, the system is unable to capture a clip or image during a routine scan.Terminated

Other IYN clearances

The newest 510(k)s for the same product code; their median review was 134 days; see every IYN device

Record

Type
Traditional, third-party review
Received
Nov 23, 2012
Review panel
Radiology
Regulation
21 CFR 892.1550

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files