Juggerknotless Soft Anchors, Juggerknotless M Soft Anchors, Juggerknotless Soft Anchors With Needles, Juggerknotless Sof
Fastener, Fixation, Nondegradable, Soft Tissue. 510(k) K123485, from Biomet Sports Medicine (Warsaw, IN). Product code MBI, class 2.
As of Nov 18, 2026, K123485 is an FDA 510(k) held by Biomet Sports Medicine, substantially equivalent on Mar 18, 2013 after 125 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 125 days
- Recalls: 2
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe JuggerKnotless™ Soft Anchors are indicated for soft tissue reattachment procedures in the shoulder, foot/ankle, elbow, knee, hand/wrist, and hip. Specific indications are as follows: Shoulder Acromio-clavicular Separation, Anterior Shoulder Instability Repair, Bankart lesion repair, Biceps tenodesis, Capsule Repair or Capsulolabral Reconstruction, Deltoid Repair, Rotator Cuff Repair, S.L.A.P…
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Biomet Sports Medicine Anchor Devices and Ziploop ConstructsK080088 · Biomet Manufacturing, Inc. · Jun 11, 2008
- Biomet Sports Medicine Juggerknot Soft AnchorK110145 · Biomet, Inc. · Mar 4, 2011
Recalls
Recalls FDA ties to K123485, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Apr 1, 2016 | Z-1828-2016Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is… | Terminated |
| Oct 10, 2013 | Z-0364-2014Investigation identified that the design of the drill bit may make the suture anchor difficult to insert into the bone. The implant may skip across the hole or miss the hole during insertion. | Terminated |
Other MBI clearances
The newest 510(k)s for the same product code; their median review was 30 days; see every MBI device- Arthrex FiberButtonK263477 · Arthrex, Inc. · Oct 27, 2026
- Q-FIX KNOTLESS All-Suture Anchor, 1.8mm, with one ULTRABRAID Suture, Blue; Q-FIX KNOTLESS All-Suture Anchor, 1.8mm,…K263348 · Smith & Nephew, Inc. · Oct 23, 2026
- TeKBrace Knotless AnchorK263352 · TheraMicro · Oct 21, 2026
- Integrity Retractable Implant Inserter (6000127)K261869 · Anika Therapeutics, Inc. · Sep 17, 2026
- Aevumed SUPERCYNCH™ Button SystemK253065 · Aevumed, Inc. · Sep 2, 2026
- ArthroTAK Tendon Anchor KitK262809 · ArthroTAK, LLC · Sep 2, 2026
- Iconix Knotless TK262677 · Riverpoint Medical · Aug 27, 2026
- Grappler Suture Anchor SystemK262047 · Paragon 28, Inc. · Aug 3, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 13, 2012
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3040
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files