Spinefrontier Pedfuse Pedicle Screw System
Orthosis, Spondylolisthesis Spinal Fixation. 510(k) K123164, from Spinefrontier, Inc. (Beverly, MA). Product code MNH, class 2.
As of Nov 17, 2026, K123164 is an FDA 510(k) held by Spinefrontier, Inc., substantially equivalent on Feb 15, 2013 after 129 days, with 6 recalls tied to it in FDA's recall database.
- Substantially equivalent: Feb 15, 2013
- Review time: 129 days
- Recalls: 6
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe SpineFrontier® PedFuse® Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- KRD1 Pedicle Screw SystemK092420 · Spinefrontier, Inc. · Nov 19, 2009
- Synergy(Tm) Ti Posterior Spinal SystemK950099 · Buckman Co., Inc. · Jan 16, 1997
- Click'XK992739 · Synthes Spine · Feb 8, 2000
- Mis SystemK100605 · Spine Wave, Inc. · Jun 4, 2010
- Modification To Cd Horizon Spinal SystemK061591 · Medtronic Sofamor Danek, Inc. · Sep 8, 2006
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- MEGAFIX® Pedicle Screw Spinal SystemK191655 · Seohancare Co., Ltd. · Mar 6, 2020
- PedFuse Pedicle Screw SystemK143377 · Spinefrontier, Inc. · Feb 12, 2015
- Spinefrontier Pedfuse Pedicle Screw SystemK133153 · Spinefrontier, Inc. · May 14, 2014
Recalls
Recalls FDA ties to K123164, newest first; each opens FDA's recall record| Mar 5, 2015 | Z-1479-2015The PedFuse Pedicle Screw Rod Caliper included with the MISquito Pedicle Screw System may be used incorrectly as the Surgical Technique does not adequately convey the intended use of the design. | Terminated |
| Nov 20, 2013 | Z-0555-2014Screw inserters may not mate properly with hex interface of the screws. | Terminated |
| Nov 20, 2013 | Z-0554-2014Screw inserters may not mate properly with hex interface of the screws. | Terminated |
| Nov 20, 2013 | Z-0553-2014Screw inserters may not mate properly with hex interface of the screws. | Terminated |
| Oct 18, 2013 | Z-0544-2014During routine servicing, it was discovered that one 8mmx55mm Reset screw was assembled with an 8mm tulip head and a 9mm screw. The implant should have consisted of an 8mm tulip head and an 8mm screw. A recall was… | Terminated |
| Sep 11, 2013 | Z-0624-2014Bone Probes used to create a pathway in pedicle bone were manufactured and marked to incorrect specifications. | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 9, 2012
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3070
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files