KRD1 Pedicle Screw System
Orthosis, Spinal Pedicle Fixation. 510(k) K092420, from Spinefrontier, Inc. (Beverly, MA). Product code MNI, class 2.
As of Nov 17, 2026, K092420 is an FDA 510(k) held by Spinefrontier, Inc., substantially equivalent on Nov 19, 2009 after 105 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Nov 19, 2009
- Review time: 105 days
- Recalls: 2
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe KRD1™ Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Synergy(Tm) Ti Posterior Spinal SystemK950099 · Buckman Co., Inc. · Jan 16, 1997
- Click'XK992739 · Synthes Spine · Feb 8, 2000
- Synthes Pangea SystemK052123 · Synthes (Usa) · Sep 23, 2005
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- PedFuse Pedicle Screw SystemK143377 · Spinefrontier, Inc. · Feb 12, 2015
- Spinefrontier Pedfuse Pedicle Screw SystemK133153 · Spinefrontier, Inc. · May 14, 2014
- Spinefrontier Pedfuse Pedicle Screw SystemK123164 · Spinefrontier, Inc. · Feb 15, 2013
- Lancer Pedicle Screw SystemK110633 · Spinal Solutions, LLC · Dec 15, 2011
Recalls
Recalls FDA ties to K092420, newest first; each opens FDA's recall record| Mar 14, 2013 | Z-1926-2013Inserters have a set screw that fixes a collar to the inserter. The screw set can loosen and fall out of the inserter which allows the collar to slide off the proximal end of the inserter. When the collar is not in… | Terminated |
| Dec 30, 2011 | Z-1737-2013One of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles. | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 6, 2009
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3070
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files