Acuson X700 Diagnstic Ultrasound System
System, Imaging, Pulsed Doppler, Ultrasonic. 510(k) K123001, from Siemens Medical Solutions USA, Inc. (Moutain View, CA). Product code IYN, class 2.
As of Nov 21, 2026, K123001 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Oct 26, 2012 after 29 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 29 days
- Recalls: 2
- Cited as a predicate by: 4 devices
The Siemens ACUSON X700 TM ultrasound imaging system is intended for the following applications: Cardiac (Adult, Pediatric), Transesophageal (Cardiac), Intracardiac, Cerebrovascular, Peripheral Vessel, Abdominal, Renal, Fetal, Abdominal, Intra-operative, Pediatric, Small Organ, Neonatal Cepahalic, Adult Cephalic, Orthopedics, Musculoskeletal Conventional, Musculoskeletal Superficial, Pelvic…
- Acuson X300 Diagnostic Ultrasound System Sonovista X300 Diagnostic Ultrasound SystemK121699 · Siemens Medical Solutions USA, Inc. · Jun 22, 2012
- Acuson S2000 Diagnostic Ultrasound SystemK112596 · Siemens Medical Solutions USA, Inc., Ultrasound DI · Sep 16, 2011
- Sonixtablet Ultrasound ScannerK113663 · Ultrasonix Medical Corporation · Jan 25, 2012
- Sonixtablet Ultrasound ScannerK102997 · Ultrasonix Medical Corporation · Dec 21, 2010
- ACUSON S1000/ S2000/ S3000 Diagnostic Ultrasound SystemsK142876 · Siemens Medical Solutions USA, Inc. · Oct 23, 2014
- SC2000 and ACUSON X300 Diagnostic Ultrasound SystemsK142395 · Siemens Medical Solutions USA, Inc. · Sep 9, 2014
- Acuson X700 Ultrasound System, Acuson X600 Ultrasound SystemK141846 · Siemens Medical Solutions USA, Inc. · Aug 7, 2014
- Acuson S1000 / S2000 / S3000K132804 · Siemens Medical Solutions USA, Inc. Ultrasound Gro · Oct 24, 2013
| Sep 5, 2015 | Z-1572-2016Due to a communication error between the software and V5Ms transducers rotation function, ACUSON X700 ultrasound systems at software versions 1.1.04, display an IMG_15 error message causes the system to lock up, which… | Terminated |
| Apr 2, 2014 | Z-1504-2014When fourSight 4D ultrasound imaging, 3-Scrape real time 3D imaging, or Anatomical M-mode are used during an Obstetric or Gynecology study, data from the previous patient could replace the current patient's date in the… | Terminated |
- Diagnostic Ultrasound System (Nuewa A20W); Diagnostic Ultrasound System (Nuewa A20); Diagnostic Ultrasound System…K261585 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Oct 16, 2026
- Diagnostic Ultrasound System (Resona I10W);Diagnostic Ultrasound System (Resona I10 Exp);Diagnostic Ultrasound System…K261617 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Oct 8, 2026
- V10 Diagnostic Ultrasound System; cV10 Diagnostic Ultrasound System; H10 Diagnostic Ultrasound System; XV10 Diagnostic…K263150 · Samsung Medison Co., Ltd. · Oct 8, 2026
- Voluson Expert 18; Voluson Expert 20; Voluson Expert 22K261933 · GE Medical Systems Ultrasound and Primary care Diagnostics, LLC · Oct 2, 2026
- EPIQ Inspira Series Diagnostic Ultrasound System; Affiniti Series Diagnostic Ultrasound SystemK262548 · Philips Ultrasound, LLC · Sep 25, 2026
- EVO X30 Diagnostic Ultrasound System; EVO X20 Diagnostic Ultrasound System; EVO X10 Diagnostic Ultrasound System; EVO…K263182 · Samsung Medison Co., Ltd. · Sep 25, 2026
- Vivid Explorer/Vivid Advanced/Vivid FocusK261313 · GE Medical Systems Ultrasound and Primary Care Diagnostics · Sep 2, 2026
- Digital Color Doppler Ultrasound System (Marvo 95/Marvo 95 Elite/Marvo 90/Marvo 90 CV/Marvo 90 Pro/Marvo 90 Exp/Marvo…K254027 · Sonoscape Medical Corp. · Aug 24, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Sep 27, 2012
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1550
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files