GE Innova/Innova Igs/Discovery Igs/Optima Angiographic, Fluoscopic X-Ray Systems With Cathlab Frontiers Solutions

Interventional Fluoroscopic X-Ray System. 510(k) K122457, from GE Healthcare (Buc, FR). Product code OWB, class 2.

As of Nov 19, 2026, K122457 is an FDA 510(k) held by GE Healthcare, substantially equivalent on Jan 2, 2013 after 142 days, with 20 recalls tied to it in FDA's recall database.

  • Review time: 142 days
  • Recalls: 20
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K122457, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 15, 2019Z-1200-2019There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.Open
Mar 29, 2018Z-2387-2018, Z-2386-2018The Discovery IGS and Innova IGS systems may experience X-ray abort errors which could result in the loss of imaging mode before or during fluoroscopic use.Terminated
Dec 18, 2015Z-0763-2016GE Healthcare has recently become aware of a reported incident in which a Mavig suspension arm with monitor assembly fell to the floor.Terminated
Nov 13, 2015Z-0413-2016GE Healthcare has recently become aware of a potential safety issue related to a non-recoverable loss of displayed imaging (loss of monitor video) involving Optima CL323i & Optima IGS 320 systems .Terminated
Nov 13, 2015Z-0414-2016GE Healthcare has recently become aware of a potential safety issue related to a non-recoverable loss of displayed imaging (loss of monitor video) involving Optima CL323i & Optima IGS 320 systems.Terminated
May 29, 2015Z-1979-2015, Z-1977-2015, Z-1974-2015, Z-1982-2015, Z-1985-2015, Z-1980-2015, Z-1973-2015, Z-1978-2015, Z-1984-2015, Z-1981-2015, Z-1976-2015, Z-1983-2015GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery).Terminated
Sep 4, 2014Z-0451-2015Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.Terminated
Nov 15, 2013Z-0853-2014GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems. While performing fluoroscopy on a biplane fluoroscopy unit, there is a potential for loss of the x-ray imaging…Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Aug 13, 2012
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files