Neuviz 64 Multi-Slice Ct Scanner System

System, X-Ray, Tomography, Computed. 510(k) K121792, from Neusoft Medical Systems Co., Ltd. (Echo, OR). Product code JAK, class 2.

As of Nov 17, 2026, K121792 is an FDA 510(k) held by Neusoft Medical Systems Co., Ltd., substantially equivalent on Nov 16, 2012 after 141 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Nov 16, 2012
  • Review time: 141 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The NeuViz 64 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate
  • ClearviewK133373 · Neusoft Medical Systems Co., Ltd. · Dec 3, 2014

Recalls

Recalls FDA ties to K121792, newest first; each opens FDA's recall record
Oct 22, 2018Z-1279-2019Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the…Terminated
Aug 31, 2016Z-0876-2017The Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be interrupted. The system must be powered down and restarted to clear the…Terminated
Aug 31, 2016Z-0875-2017The Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be interrupted. The system must be powered down and restarted to clear the…Terminated
May 2, 2015Z-1650-2016It was found by R&D in April, 2014 that there was a defect in NeuViz 64 system software version 1.0.5+P09. The icons of side decubitus didn't meet the patient position description, when the scanner position was set to…Terminated
Mar 8, 2014Z-1693-2015The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it. The thinnest part of the compensator is subject to radiation degradation and as a result has been found to…Terminated

Other JAK Clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Record

Type
Traditional
Received
Jun 28, 2012
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files