Neuviz 16 Multi-Slice Ct Scanner System

System, X-Ray, Tomography, Computed. 510(k) K092742, from Neusoft Medical Systems Co., Ltd. (Shenyang, Liaoning, CN). Product code JAK, class 2.

As of Nov 17, 2026, K092742 is an FDA 510(k) held by Neusoft Medical Systems Co., Ltd., substantially equivalent on Sep 22, 2009 after 14 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 22, 2009
  • Review time: 14 days
  • Recalls: 5
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The NeuViz 16 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multislice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional imagers of the body from the same axial plane taken at different angles.

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K092742, newest first; each opens FDA's recall record
Mar 24, 2016Z-1573-2016The following issues are found in NeuViz 16 systems with software version 1.1.4.21425 and version 1.1.4.21426: 1) During the filming operation on MX 16-slice console software, the clipboard used for copying and pasting…Terminated
Mar 4, 2013Z-1165-2013User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the…Terminated
Oct 18, 2012Z-0466-2013The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.Terminated
Jun 8, 2012Z-2083-2012One of the four screws that secure the Oil-accumulator to the X-Ray Tube pulled out in the NeuViz 16 CT scanner, which caused the Oil-accumulator to detach.Terminated
Feb 9, 2011Z-2241-2011Dimension measurement displays on a combined image were found inaccurate after the clinician merged multiple image series into one image.Terminated

Other JAK Clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Record

Type
Traditional, third-party review
Received
Sep 8, 2009
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files