Fresenius 2008T Hemodialysis Machine With Bibag System
Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K121341, from Fresenius Medical Care North America (Waltham, MA). Product code KDI, class 2.
As of Nov 20, 2026, K121341 is an FDA 510(k) held by Fresenius Medical Care North America, substantially equivalent on Dec 6, 2012 after 217 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 217 days
- Recalls: 5
- Cited as a predicate by: 3 devices
The modified Fresenius 2008T Hemodialysis Machine with bibag System has the same indications for use as the unmodified device. The Fresenius Medical Care bibag system is used with Fresenius Medical Care three stream proportioning hemodialysis systems equipped with the bibag module such as the 2008T Hemodialysis Machine and is intended for use in bicarbonate hemodialysis for acute and chronic…
- Fresenius 2008 Hemodiaysis Machine With Bibag SystemK120017 · Fresenius Medical Care North America · Feb 2, 2012
- Fresenius 2008T Hemodialysis MachineK120505 · Fresenius Medical Care North America, Design Cente · Mar 6, 2012
- Acetic Acid Wand, Acid Wand Assembly, Bicarbonate Wand AssemblyK133299 · Fresenius Medical Care, North America - Renal Ther · Jul 7, 2014
- 2008T Hemodialysis Machine With Optional Crit-Line Monitor (Clic)K131908 · Fresenius Medical Care Renal Therapies Group, LLC · Oct 11, 2013
- 2008K@HOME Hemodialysis MachineK124035 · Fresenius Medical Care North America · Jul 3, 2013
| Oct 5, 2017 | Z-0829-2018While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate.… | Terminated |
| Mar 10, 2016 | Z-1325-2016Added Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12) | Terminated |
| Feb 20, 2015 | Z-1550-2015The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen. | Terminated |
| Dec 16, 2014 | Z-1015-2015Unqualified power assemblies shipped to customers. | Terminated |
| Apr 25, 2014 | Z-1618-20142008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pump | Terminated |
- multiFiltratePRO SystemK261203 · Fresenius Medical Care Renal Therapies Group, LLC · Oct 19, 2026
- Careflux H60;Careflux H70;Careflux V500K261267 · Fresenius Medical Care Renal Therapies Group, LLC · Jul 24, 2026
- NxStage System One with NxViewK252377 · Fresenius Medical Care Renal Therapies Group, LLC · May 15, 2026
- Elisio™-HK260533 · Nipro Medical Corporation · Mar 19, 2026
- multiFlux 130 (F00013123); multiFlux 160 (F00013124)K252459 · Fresenius Medical Care Renal Therapies Group, LLC · Feb 10, 2026
- Tablo Hemodialysis System (PN-0008000, PN-0006000U)K253412 · Outset Medical, Inc. · Jan 26, 2026
- FX CorAL 40; FX CorAL 50K253518 · Fresenius Medical Care Renal Therapies Group, LLC · Dec 18, 2025
- Purema H Hemoconcentrator - PediatricK243920 · Medica USA, Inc. · Sep 19, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 3, 2012
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.5860
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files