Fresenius 2008 Hemodiaysis Machine With Bibag System

Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K120017, from Fresenius Medical Care North America (Waltham, MA). Product code KDI, class 2.

As of Nov 20, 2026, K120017 is an FDA 510(k) held by Fresenius Medical Care North America, substantially equivalent on Feb 2, 2012 after 30 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 30 days
  • Recalls: 3
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The modified Fresenius 2008T Hemodialysis Machine with bibag System has the same indications for use as the unmodified device. The Fresenius Medical Care bibag System is used with Fresenius Medical Care three strøam proportioning hemodialysis systems equipped with the bibag module such as the 2008T Hemodialysis Machine and is intended *Page 131* *K120017 page 4 of 6* *Fresenius Medical Care* *V*…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K120017, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 5, 2017Z-0829-2018While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate.…Terminated
Mar 10, 2016Z-1325-2016Added Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12)Terminated
Feb 20, 2015Z-1550-2015The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.Terminated

Other KDI clearances

The newest 510(k)s for the same product code; their median review was 189 days; see every KDI device

Record

Type
Special
Received
Jan 3, 2012
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files