Vivid I Diagnostic Ultrasound System, Vivid Q Diagnostic Ultrasound System
Transducer, Ultrasonic, Diagnostic. 510(k) K121062, from Ge Medical System Israel , Ltd. (Wauwatosa, WI). Product code ITX, class 2.
As of Nov 20, 2026, K121062 is an FDA 510(k) held by Ge Medical System Israel , Ltd., substantially equivalent on Aug 17, 2012 after 133 days, with 1 recall tied to it in FDA's recall database.
- Review time: 133 days
- Recalls: 1
- Cited as a predicate by: 3 devices
The Vivid i/q ultrasound systems are intended for use by, or under the direction of, a qualified physician for ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculoskeletal Conventional and Superficial; Urology (including prostate)…
- Vivid I Diagnostic Ultrasound SystemK102388 · Ge Medical System Israel , Ltd. · Nov 4, 2010
- GE Vivid E9 Diagnostic Ultrasound Imaging SystemK101149 · Ge Vingmed Ultrasound AS · Oct 13, 2010
- ACUSON P500 Ultrasound SystemK163396 · Siemens Medical Solutions USA, Inc. · Jan 4, 2017
- Vivid iqK161706 · Ge Medical Systems Ultrasound and Primary Care Diagnostics L · Aug 5, 2016
- M9/M9CV/M9T Diagnostic Ultrasound SystemK152543 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Oct 29, 2015
| Started | Recall | Status |
|---|---|---|
| Nov 10, 2022 | Z-0757-2023Potential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch fire. | Open |
- Align® Cover and Pad Guidance ToolK260114 · CIVCO Medical Instruments Co., Inc. · Sep 24, 2026
- BA-004CCK261170 · Fujifilm Corporation · Aug 12, 2026
- EM-TECH Passive Ultrasound Probe StrapK254143 · EM-TECH Solutions, Inc. · Aug 10, 2026
- SmartGuide Disposable Guides and GridsK260161 · UC-CARE , Ltd. · Jul 30, 2026
- ND Needle Guidance SystemK262474 · Accurate Access, LLC · Jul 27, 2026
- Accuro® 3S Needle Guide KitK261014 · Rivanna Medical, Inc. · Jul 24, 2026
- Ultrasonic Probe UM-3R (UM-3R); Ultrasonic Probe UM-G20-29R (UM-G20-29R)K250883 · Olympus Medical Systems Corporation · Sep 18, 2025
- Ultrasonic Probe UM-S20-17S (UM-S20-17S); Ultrasonic Probe UM-S20-20R (UM-S20-20R)K250762 · Olympus Medical Systems Corporation · Jul 11, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 6, 2012
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1570
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files