Insufflator, Laparoscopic. 510(k) K120151, from Northgate Technologies, Inc. (Elgin, IL). Product code HIF, class 2.

As of Nov 20, 2026, K120151 is an FDA 510(k) held by Northgate Technologies, Inc., substantially equivalent on Aug 17, 2012 after 212 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 212 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Nebulae™ I 50 LPM Insufflator provides CO₂ gas distention of surgical cavities for diagnostic and/or operative endoscopy. The insufflator has multiple operating modes which can be used for the following procedures: General Laparoscopic Pediatric Laparoscopic Bariatric Laparoscopic Minimally Invasive Vessel Harvesting

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K120151, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 15, 2021Z-1147-2021An internal tubing component in the flow path was made with incorrect material which was not tested for use with this device.Terminated
Nov 21, 2016Z-0866-2017Device could cause a patient overpressure situation without any visual or auditory indication or warningTerminated

Other HIF clearances

The newest 510(k)s for the same product code; their median review was 223 days; see every HIF device

Record

Type
Traditional
Received
Jan 18, 2012
Review panel
Obstetrics/Gynecology
Regulation
21 CFR 884.1730

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files