45L Core Insufflator, Model F114

Insufflator, Laparoscopic. 510(k) K063367, from W.O.M. World of Medicine AG (Cambridge, MA). Product code HIF, class 2.

As of Nov 20, 2026, K063367 is an FDA 510(k) held by W.O.M. World of Medicine AG, substantially equivalent on May 23, 2007 after 197 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 197 days
  • Recalls: 4
  • Cited as a predicate by: 7 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The 45L CORE Insufflator F114 is a CO2 insufflator intended for use in diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. The high flow application, the low flow application and the bariatric application of the device are each indicated for use in facilitating the use of a laparoscope by filling the peritoneal cavity with gas to distend it.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

7 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K063367, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 7, 2019Z-1727-2019, Z-1726-2019The heated tube sets were consistently leaking and detaching from the cassette.Terminated
Mar 28, 2016Z-1566-2016When operating the device in high flow mode (40L/min, 15 mmHg) and connecting to a house gas inlet source, within seconds of insufflating the flow rate would drop to 0L/min and the red Check Gas Supply symbol would…Terminated
Dec 29, 2008Z-1198-2010Unexpected powering down/failure to boot up: A problem with a transistor in the heater/temperature sensor component in the device may cause the device to power off during surgery or to fail to boot up at all. The risk…Terminated

Other HIF clearances

The newest 510(k)s for the same product code; their median review was 223 days; see every HIF device

Record

Type
Traditional
Received
Nov 7, 2006
Review panel
Obstetrics/Gynecology
Regulation
21 CFR 884.1730

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files