Arca
Skip to content

Zimmer Dental Tapered Screw-Vent M Implant

Implant, Endosseous, Root-Form. 510(k) K111889, from Zimmer Dental, Inc. (Carlsbad, CA). Product code DZE, class 2.

As of Nov 17, 2026, K111889 is an FDA 510(k) held by Zimmer Dental, Inc., substantially equivalent on Dec 6, 2011 after 158 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Dec 6, 2011
  • Review time: 158 days
  • Recalls: 1
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Tapered Screw-Vent® M Implants are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K111889, newest first; each opens FDA's recall record
StartedReasonStatus
May 10, 2018Z-2963-2018Dental implants have incorrect item/lot numbers and size on the inner package labeling. Correct item/lot numbers and size are on the outer package labeling.Terminated

Other DZE Clearances

The newest 510(k)s for the same product code; their median review was 257 days; see every DZE device

Record

Type
Traditional
Received
Jul 1, 2011
Review panel
Dental
Regulation
21 CFR 872.3640

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files