System, X-Ray, Fluoroscopic, Image-Intensified. 510(k) K111292, from Siemens Medical Solutions USA, Inc. (Mavern, PA). Product code JAA, class 2.

As of Nov 18, 2026, K111292 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Jun 3, 2011 after 28 days, with 6 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jun 3, 2011
  • Review time: 28 days
  • Recalls: 6
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Luminos Agile is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using either film cassettes or a digital mobile flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. Luminos Agile may be used for emergency treatment on an outpatient basis, as well as for bedside examinations.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K111292, newest first; each opens FDA's recall record
Dec 9, 2022Z-1001-2023There is potential for the footrest to detach from the patient table during use.Open, Classified
Apr 13, 2015Z-1669-2015Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing…Terminated
Dec 15, 2014Z-0994-2015Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be…Terminated
Sep 2, 2014Z-0005-2015Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause…Terminated
Sep 3, 2013Z-0374-2014A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required…Terminated
Feb 15, 2012Z-1159-2012Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images…Terminated

Other JAA Clearances

The newest 510(k)s for the same product code; their median review was 110 days; see every JAA device

Record

Type
Special
Received
May 6, 2011
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.