Shunt, Central Nervous System And Components. 510(k) K110206, from Aesculap, Inc. (Center Valley, PA). Product code JXG, class 2.

As of Nov 20, 2026, K110206 is an FDA 510(k) held by Aesculap, Inc., substantially equivalent on Oct 27, 2011 after 275 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 275 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

INDICATIONS FOR USE The Mithke Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K110206, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 11, 2014Z-0814-2015Aesculap Inc. (AIC (USA)) initiated a recall of Miethke Shunt System accessories due to missing patients' labels for the miniNAV (FV660T). There are no patient injuries resulting from this issue.Terminated
Jun 4, 2012Z-0805-2013The deflector may not move freely on the ventricular catheter. The deflector comes pre-positioned at the 5 cm mark on the catheter. In some cases the surgeon will want to adjust the deflector to a different location on…Terminated

Other JXG clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every JXG device

Record

Type
Traditional
Received
Jan 25, 2011
Review panel
Neurology
Regulation
21 CFR 882.5550

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.