Shunt, Central Nervous System And Components. 510(k) K042465, from Medtronic Neurosurgery (Goleta, CA). Product code JXG, class 2.

As of Nov 19, 2026, K042465 is an FDA 510(k) held by Medtronic Neurosurgery, substantially equivalent on Oct 19, 2004 after 36 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 36 days
  • Recalls: 4
  • Cited as a predicate by: 6 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Medtronic PS Medical Strata™ II Valve and shunt assemblies with and without BioGlide are shunt components designed to provide continued CSF flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K042465, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 11, 2021Z-1116-2021, Z-1115-2021Potential for variation in radiopaque marking visibility under radiographic imaging for adjustable and fixed-pressure valves that are used in the management of hydrocephalusOpen
Jan 17, 2017Z-1121-2017Medtronic Neurosurgery initiated a recall after receiving five (5) Product Experiences between February 15, 2015 and August 23, 2016 reporting of discrepancies between the pressure/performance level (PL) at which the…Terminated
Nov 18, 2013Z-0399-2014Medtronic Neurosurgery is recalling the CSF-Unitized Shunts, Delta Shunt Assembly Kit, and the Strata II Shunt Assembly Kit because they were found to have label discrepancies.Terminated

Other JXG clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every JXG device

Record

Type
Special
Received
Sep 13, 2004
Review panel
Neurology
Regulation
21 CFR 882.5550

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files