Achieva R4 1.5T And Achieva R4 3.0T

System, Nuclear Magnetic Resonance Imaging. 510(k) K110151, from Philips Health Care (Bothell, WA). Product code LNH, class 2.

As of Nov 18, 2026, K110151 is an FDA 510(k) held by Philips Health Care, substantially equivalent on Mar 22, 2011 after 63 days, with 20 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Mar 22, 2011
  • Review time: 63 days
  • Recalls: 20
  • Cited as a predicate by: 8 devices

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

8 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K110151, newest first; each opens FDA's recall record
Aug 28, 2026Z-3314-2026When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction…Open, Classified
Aug 21, 2026Z-3243-2026A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.Open, Classified
Aug 11, 2021Z-2411-2021When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatmentCompleted
Jul 30, 2021Z-2346-2021Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injuryTerminated
Dec 21, 2018Z-0924-2019There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.Terminated
Dec 21, 2018Z-0926-2019There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.Terminated
Dec 21, 2018Z-0925-2019There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.Terminated
Dec 1, 2016Z-0825-2017Possibility that a patient or user may be exposed to a very small level of touch current.Terminated
Aug 21, 2014Z-0325-2016When a fused series of a sagittal, coronal or radial multi station scan is generated in MobiView, the resultant image order is reversed. This occurs with software release R5.1.7/R5.1.8.Terminated
Jul 27, 2014Z-2237-2014The following software issues have been identified in the affected products. Problem 1: When reopening a bookmark generated from processing a MUGA (Multi-Gated Acquisition) scan within the NM Cardiac Application, the…Terminated
Jun 13, 2014Z-1863-2014For some identified Ingenia systems shipped between December 2013 and March 2014, cover set rear magnetic M10 nuts were erroneously supplied instead of A4 stainless steel nuts.Terminated
Jun 13, 2014Z-1862-2014For some identified Ingenia systems shipped between December 2013 and March 2014, cover set rear magnetic M10 nuts were erroneously supplied instead of A4 stainless steel nuts.Terminated
Jun 13, 2014Z-1864-2014For some identified Ingenia systems shipped between December 2013 and March 2014, cover set rear magnetic M10 nuts were erroneously supplied instead of A4 stainless steel nuts.Terminated
Jun 13, 2014Z-1865-2014For some identified Ingenia systems shipped between December 2013 and March 2014, cover set rear magnetic M10 nuts were erroneously supplied instead of A4 stainless steel nuts.Terminated
May 15, 2014Z-1827-2014Ingenia customers have experienced clamping of the foot under the central column of the Height Adjustable Flex Trak trolley and the Trolley Variable Height IRF.Terminated
May 15, 2014Z-1826-2014Ingenia customers have experienced clamping of the foot under the central column of the Height Adjustable Flex Trak trolley and the Trolley Variable Height IRF.Terminated
May 5, 2014Z-0135-2016In spine clinical workflows, cross reference lines may be used to determine the position of slices. In cases, where MobiView fused Images are used to show the cross reference lines, the cross reference lines may be…Terminated
Mar 11, 2014Z-1678-2014During SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. In addition, resolution recovery is not…Terminated
Jan 27, 2012Z-1125-2012There maybe incorrect values in the diffusion/perfusion maps.Terminated
Aug 3, 2011Z-0293-2012The firm determined that during scanning procedures, for some larger patients, some reddening of the skin of can occur in certain instances due to RF heating when the torso of the patient is positioned within the body…Terminated

Other LNH Clearances

The newest 510(k)s for the same product code; their median review was 152 days; see every LNH device

Record

Type
Traditional
Received
Jan 18, 2011
Review panel
Radiology
Regulation
21 CFR 892.1000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files