Panorama 1.0T

System, Nuclear Magnetic Resonance Imaging. 510(k) K041602, from Philips Medical Systems North America, Inc. (Bothell, WA). Product code LNH, class 2.

As of Nov 19, 2026, K041602 is an FDA 510(k) held by Philips Medical Systems North America, Inc., substantially equivalent on Jul 26, 2004 after 42 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 42 days
  • Recalls: 7
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

PANORAMA 1.0T is a Magnetic Resonance Diagnostic Device intended for general diagnostic use that produces transverse, sagittal, coronal and oblique cross-sectional images based upon ^1H metabolites, and that displays the internal structure of the whole body. These images when interpreted by a trained physician, yield information that may assist in diagnosis. *XJR-148-6269/bf* *Page 2 of 2*

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K041602, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 17, 2023Z-0505-2024Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain…Open
Dec 21, 2018Z-0898-2019, Z-0917-2019There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.Terminated
Mar 16, 2018Z-1434-2018Potential risk for helium gas inside the MR examination room during a magnet quenchTerminated
Feb 5, 2010Z-1458-2011In the ExamCard MobiFlex the Patient Position is "Prone" which must be "Supine". The result is mislabeling of left and right side of the patient. If this is not detected by the operator there is a risk of wrong…Terminated
May 12, 2009Z-0794-2011The Accessory Cart contained magnetic material and is attracted to the magnet when bringing it close to the system.Terminated
Apr 30, 2007Z-0984-2007Potential for gaseous helium to improperly vent, resulting in rupture of the magnet vessel and venting of helium into the MRI site.Terminated

Other LNH clearances

The newest 510(k)s for the same product code; their median review was 143 days; see every LNH device

Record

Type
Traditional
Received
Jun 14, 2004
Review panel
Radiology
Regulation
21 CFR 892.1000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files