Quadrox-I Pediatric Microporous Membrane Oxygenator With Integrated Arterial Filter With Softline / Bioline Coating
Oxygenator, Cardiopulmonary Bypass. 510(k) K103191, from Maquet Cardiopulmonary, AG (Hirrlingen, DE). Product code DTZ, class 2.
As of Nov 19, 2026, K103191 is an FDA 510(k) held by Maquet Cardiopulmonary, AG, substantially equivalent on Mar 8, 2011 after 131 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 131 days
- Recalls: 4
- Cited as a predicate by: 2 devices
The membrane oxygenator QUADROX-i Pediatric is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. It is designed for a blood flow rate of 0.2 – 2.8 l/min and is intended for the treatment of pediatric patients. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide and regulates the blood temperature.
- D101 Kids Infant Hollow Fiber Membrane Oxygenator With Integrated Hardshell Cardiotomy/Venous ReservoirK072091 · Sorin Group Italia S.R.L. · Oct 29, 2007
- Capiox Arterial FilterK943917 · Terumo Medical Corp. · Jul 25, 1995
- Quadrox-I, Quadrox-Id, Quadrox-Ir OxygenatorsK132829 · Maquet Cardiopulmonary, AG · Oct 8, 2013
- Quadrox-I Microporporous Membrane Oxgenator Series and Qudrox-Id Diffusion Membrane Oxygenators SeriesK112360 · Maquet Cardiopulmonary, AG · Sep 15, 2011
| May 18, 2023 | Z-1655-2023Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis,… | Open |
| May 18, 2023 | Z-1641-20231. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may… | Open |
| Aug 15, 2013 | Z-2262-2013The manufacturer has conducted a review of the results of product testing performed during the last twelve months and has identified specific lots of product which may not meet specifications for bond strength on the… | Terminated |
| Jun 12, 2012 | Z-0160-2013It has come to firm's attention that, in some rare cases, the blood outlet connector on some models of MAQUET Oxygenators may detach from the oxygenator. This event has occurred in about 0.01% (1:10,000) of units… | Terminated |
- HORIZON 8AF (Integrated OXY MODULE 8AF and Horizon CVR) (US5305)K261438 · Eurosets S.R.L · Sep 14, 2026
- Lifemotion Disposable Membrane OxygenatorK253838 · Lifemotion Medical Technology Co., Ltd. · Mar 16, 2026
- Quantum Perfusion Blood Oxygenator ECC VT160-E3 (VT160-E3); Quantum Perfusion Blood Oxygenator ECC VT200-E3 (VT200-E3);…K240934 · Spectrum Medical S.R.L. · Mar 12, 2026
- Horizon AF Plus System (Integrated A.L. One AF Plus OXYGENATOR and Horizon CVR) (US5300); A.L. One AF Plus OXYGENATOR…K250821 · Eurosets S.R.L · Dec 12, 2025
- Trilly Pediatric Oxygenator with Integrated Arterial Filter (US5232)K242503 · Eurosets S.R.L · Nov 24, 2025
- Quantum Perfusion Blood Oxygenator VT75-E1, Quantum Perfusion Blood Oxygenator VT75-E2K231982 · Qura S.R.L · Sep 12, 2024
- Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (BB811); Affinity Fusion Oxygenator…K240666 · Medtronic Perfusion Systems · Jul 10, 2024
- Affinity NT Oxygenator and Uncoated Cardiotomy/ Venous Reservoir (541); Affinity NT Oxygenator with Balance Biosurface…K241352 · Medtronic, Inc. · Jun 12, 2024
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 28, 2010
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.4350
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files