Somatom Definition As
System, X-Ray, Tomography, Computed. 510(k) K103127, from Siemens Medical Solutions USA, Inc. (Mavern, PA). Product code JAK, class 2.
As of Nov 17, 2026, K103127 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Mar 4, 2011 after 133 days, with 6 recalls tied to it in FDA's recall database.
- Substantially equivalent: Mar 4, 2011
- Review time: 133 days
- Recalls: 6
- Cited as a predicate by: 2 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Siemens SOMATOM Definition AS Open systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Somatom Definition, Model As/As+K081022 · Siemens Medical Solutions USA, Inc. · Jun 2, 2008
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- CT6485, CT12885K172058 · Analogic Corporation · Feb 13, 2018
- Somatom Definition As OpenK130901 · Siemens Medical Systems, Inc. · Jan 2, 2014
Recalls
Recalls FDA ties to K103127, newest first; each opens FDA's recall record| Nov 24, 2020 | Z-0771-2021System does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan | Terminated |
| Jun 22, 2020 | Z-3025-2020Sporadic problems with the current software may result in scanning workflow interruptions and unexpected user notifications causing delay in diagnosis or patient rescan | Terminated |
| Oct 7, 2019 | Z-0406-2020Software issue identified in the software versions syngo CT VB20 running on the SOMATOM CT Scanner: scan aborts and system crashes | Terminated |
| Jul 3, 2019 | Z-2268-2019Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort. | Terminated |
| Feb 14, 2019 | Z-1247-2019Potential risk of scans being aborted when using the optional Dual Spiral Dual Engery mode due to a software issue found in syngo CT VB10A for the Siemens Healthineers CT scanners with a single X-ray tube. | Terminated |
| Nov 15, 2013 | Z-0907-2014During clinical operation, a malfunction may occur causing a system display freeze with SOMATOM Definition, Definition AS, and definition Edge while running software version syngo CT2012 B and syngo CT2013B. | Terminated |
Other JAK Clearances
The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device- AVIEW Aorta MetricsK260391 · Coreline Soft Co., Ltd. · Oct 16, 2026
- CAS-One IRK261225 · Cascination AG · Sep 29, 2026
- BodyTom 64 (NL4100); OmniTom Elite (NL5000)K261125 · Samsung HME America, Inc., dba NeuroLogica Corp. · Sep 22, 2026
- SCENARIA View Phase 6.0K261229 · Fujifilm Corporation · Sep 18, 2026
- FCT iStream Phase 2K253836 · Fujifilm Corporation · Sep 4, 2026
- BodyTom Elite (NL4000)K253450 · NeuroLogica Corp., a Samsung Company · Jul 28, 2026
- CardIQ SuiteK253770 · GE Medical Systems SCS · Jul 21, 2026
- Enhanced Boundary for PCCTK253560 · Ge Medical Systems, LLC · Jun 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 22, 2010
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files