Arca
Skip to content

Somatom Definition, Model As/As+

System, X-Ray, Tomography, Computed. 510(k) K081022, from Siemens Medical Solutions USA, Inc. (Malvern, PA). Product code JAK, class 2.

As of Nov 17, 2026, K081022 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Jun 2, 2008 after 53 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jun 2, 2008
  • Review time: 53 days
  • Recalls: 3
  • Cited as a predicate by: 4 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The SOMATOM Definition AS/ AS + is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K081022, newest first; each opens FDA's recall record
May 29, 2009Z-1675-2009Support mechanism may not engage properly due to improper stowage after opening which may result in internal component damage or potential injury.Terminated
May 1, 2009Z-1442-2009Improper installation of accessory support arm, which could potentially result in the loosening of the arm over time causing it to fall.Terminated
Dec 9, 2008Z-0929-2009possible improper installation of tilt transport lock/tilt limiterTerminated

Other JAK Clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Record

Type
Traditional
Received
Apr 10, 2008
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.