Arca
Skip to content
Tools
Tools

Elitech Clinical Systems Selectra Prom/Elical 2/Elitrol I and Elitrol Ii/Ast/Got 4+1 Sl/Potassium, Sodium, Chloride...

Electrode, Ion Specific, Sodium. 510(k) K102647, from Elitechgroup Vital Scientific BV (Bothell, WA). Product code JGS, class 2.

As of Nov 19, 2026, K102647 is an FDA 510(k) held by Elitechgroup Vital Scientific BV, substantially equivalent on Mar 24, 2011 after 191 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 191 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Selectra ProM analyzer: The ELITech Clinical system Selectra ProM is an automated clinical chemistry system intended for use in clinical laboratories. It is intended to be used for a variety of assay methods that have been applied to spectrophotometric and electrochemical techniques.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K102647, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 12, 2016Z-2764-2019Software malfunction; The action being taken due to the device becoming inoperable due to software lockup.Terminated
Aug 11, 2014Z-0006-2015Some users of ENVOY 500 AST Reagent Kit, reference 55255, for Envoy 500 systems are observing that Quality Control (QC) fails before the product expiry is reached.Terminated
Jan 10, 2012Z-0158-2020, Z-0157-2020The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.Terminated

Other JGS clearances

The newest 510(k)s for the same product code; their median review was 268 days; see every JGS device

Record

Type
Traditional
Received
Sep 14, 2010
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1665

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files