Abx Pentra 400 Clinical Chemistry Analyzer (Option: I.S.E. Module)
Hexokinase, Glucose. 510(k) K052007, from Horiba Abx (Montpellier, FR). Product code CFR, class 2.
As of Nov 18, 2026, K052007 is an FDA 510(k) held by Horiba Abx, substantially equivalent on Dec 16, 2005 after 144 days, with 6 recalls tied to it in FDA's recall database.
- Review time: 144 days
- Recalls: 6
- Cited as a predicate by: 14 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe ABX PENTRA 400 is a discrete photometric benchtop chemistry analyzer for clinical use. clinical use. The device is intended to duplicate manual analytical procedures by performing various steps such as pipetting, mixing, heating and measuring color intensity. The device is intended for use in conjunction with certain materials to measure a variety of analytes.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Roche Cobas Mira PlusK920402 · Roche Diagnostic Systems, Inc. · Apr 29, 1992
- (Modified Version) Integra Ion Selective Electrode (Ise) ModuleK963627 · Roche Diagnostic Systems, Inc. · Oct 24, 1996
- Cobas Reagent for Glucose RapidK801297 · Roche Analytical Instruments, Inc. · Jun 30, 1980
- Modification To Calibrator for Automatic Systems (Cfas)K033501 · Roche Diagnostics Corp. · Nov 20, 2003
Cited as a predicate
14 later 510(k) summaries name this device as a predicate; the 12 newest- Abx Pentra Micro Alb Control L/HK133676 · Horiba Abx SAS · Jul 11, 2014
- Abx Pentra Calcium As CP, Abx Pentra Urine Control L/H, Abx Pentra Multical, Abx Pentra N Control, and Abx Pentra P ContK123171 · Horiba Abx SAS · Jul 3, 2013
- Abx Pentra Creatinine 120 Cp; Abx Pentra Multical; Abx Pentra N Control; Abx Pentra P Control; Abx Pentra Urine ControlK110530 · Horiba Abx SAS · Apr 30, 2012
- Abx Pentra Enzymatic Creatinine CP, Abx Pentra Multical, Abx Pentra N Control, Abx Pentra P Control, and Abx Pentra UrinK110137 · Horiba Abx SAS · Aug 10, 2011
- Pentra C200, Ise Module and Abx Pentra Glucose HK CP ReagentK103788 · Horiba Abx SAS · Nov 8, 2011
- Elitech Clinical Systems Selectra Prom/Elical 2/ELITROL I and Elitrol Ii/Ast/Got 4+1 Sl/Potassium, Sodium, Chloride...K102647 · Elitechgroup Vital Scientific BV · Mar 24, 2011
- Elitech Clinical Systems Glucose Pap SL Elitech Clinical Systems Elical 2, Elitech Clinical Systems Elitrol 1 and 2K100525 · Seppim S.A.S. · Dec 15, 2010
- Abx Pentra Tpu Cal, Model A11A01898K092570 · Horiba Abx SAS · Oct 16, 2009
- Abx Pentra Creatinine 120 CP, Total Protein 100 CP, Multical, N and P Controls, and Urine Control L/HK072115 · Horiba Abx · Nov 30, 2007
- Abx Pentra Urine CalK071779 · Horiba Abx · Mar 5, 2008
- Abx Pentra Glucose HK CP, Urea CP, Uric Acid CPK070146 · Horiba Abx · Oct 12, 2007
- Abx Pentra Amylase Cp; Multical; N Control; P Control; Clean-Chem Cp; Clean-Chem 99 CPK062180 · Horiba Abx · Oct 2, 2006
Recalls
Recalls FDA ties to K052007, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Nov 21, 2017 | Z-0493-2018, Z-0494-2018The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being… | Terminated |
| Jan 11, 2016 | Z-0740-2016N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce… | Terminated |
| Nov 13, 2014 | Z-0902-2015HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being… | Terminated |
| Nov 21, 2013 | Z-0904-2014Horiba Medical is recalling the ABX PENTRA N Control because it has been updated to include a revised Target Value assignment for Potassium. | Terminated |
| Mar 20, 2013 | Z-1141-2013HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All… | Terminated |
Other CFR clearances
The newest 510(k)s for the same product code; their median review was 111 days; see every CFR device- Glucose HK Gen.3; ONLINE DAT Methadone II; cobas pro integrated solutionsK253490 · Roche Diagnostics · Feb 12, 2026
- Glucose2K252357 · Abbott Ireland · Oct 24, 2025
- cobas pure integrated solutions, Glucose HK Gen.3, ISE indirect Na for Gen.2, ONLINE DAT Methadone II, Elecsys TSHK220134 · Roche Diagnostics · Sep 16, 2022
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 25, 2005
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1345
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files