Abx Pentra 400 Clinical Chemistry Analyzer (Option: I.S.E. Module)

Hexokinase, Glucose. 510(k) K052007, from Horiba Abx (Montpellier, FR). Product code CFR, class 2.

As of Nov 18, 2026, K052007 is an FDA 510(k) held by Horiba Abx, substantially equivalent on Dec 16, 2005 after 144 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 144 days
  • Recalls: 6
  • Cited as a predicate by: 14 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The ABX PENTRA 400 is a discrete photometric benchtop chemistry analyzer for clinical use. clinical use. The device is intended to duplicate manual analytical procedures by performing various steps such as pipetting, mixing, heating and measuring color intensity. The device is intended for use in conjunction with certain materials to measure a variety of analytes.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

14 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to K052007, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 21, 2017Z-0493-2018, Z-0494-2018The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being…Terminated
Jan 11, 2016Z-0740-2016N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce…Terminated
Nov 13, 2014Z-0902-2015HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being…Terminated
Nov 21, 2013Z-0904-2014Horiba Medical is recalling the ABX PENTRA N Control because it has been updated to include a revised Target Value assignment for Potassium.Terminated
Mar 20, 2013Z-1141-2013HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All…Terminated

Other CFR clearances

The newest 510(k)s for the same product code; their median review was 111 days; see every CFR device

Record

Type
Traditional
Received
Jul 25, 2005
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1345

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files