Arca
Skip to content
Tools
Tools

Mahurkar Triple Lumen Dialysis Catheter

Catheter, Hemodialysis, Triple Lumen, Non-Implanted. 510(k) K102605, from Covidien, LLC, Formerly Registered AS Tyco Healthca (Mansfield, MA). Product code NIE, class 2.

As of Nov 19, 2026, K102605 is an FDA 510(k) held by Covidien, LLC, Formerly Registered AS Tyco Healthca, substantially equivalent on Dec 22, 2010 after 103 days, with 25 recalls tied to it in FDA's recall database.

  • Review time: 103 days
  • Recalls: 25+
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The MAHURKART™ Triple Lumen Catheter is intended for short term central venous access for hemodialysis, apheresis, infusion, central venous pressure monitoring and pressure injection of contrast media. The maximum recommended infusion rate is 5 mL/sec for power injection of contrast media.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K102605, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Jun 28, 2023Z-2377-2023, Z-2371-2023, Z-2378-2023, Z-2380-2023, Z-2390-2023, Z-2379-2023, Z-2376-2023, Z-2368-2023, Z-2372-2023, Z-2385-2023, Z-2388-2023, Z-2391-2023, Z-2367-2023, Z-2381-2023, Z-2383-2023, Z-2375-2023, Z-2374-2023, Z-2387-2023, Z-2382-2023, Z-2384-2023, Z-2369-2023, Z-2389-2023, Z-2370-2023, Z-2386-2023, Z-2373-2023Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully,…Open

Other NIE clearances

The newest 510(k)s for the same product code; see every NIE device

Record

Type
Traditional
Received
Sep 10, 2010
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5540

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files