Mahurkar Triple Lumen Catheter, 12 FR
Catheter, Hemodialysis, Triple Lumen, Non-Implanted. 510(k) K020089, from The Kendall Company (Mansfield, MA). Product code NIE, class 2.
As of Nov 20, 2026, K020089 is an FDA 510(k) held by The Kendall Company, substantially equivalent on Apr 10, 2002 after 90 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 90 days
- Recalls: 25+
- Cited as a predicate by: 4 devices
Mahurkar® Triple Lumen Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion.
- Mahurkar Acute Single Lumen Catheter, Mahurkar Acute Dual Lumen Catheter, Mahurkar Acute Triple Lumen Catheter,…K192302 · Covidien, LLC · Jan 17, 2020
- Mahurkar Triple Lumen Dialysis CatheterK102605 · Covidien, LLC, Formerly Registered AS Tyco Healthca · Dec 22, 2010
- Power-Trialysis Triple Lumen Acute Dialysis CatheterK083675 · C.R. Bard, Inc. · Mar 19, 2009
- Medcomp T-3K033570 · Medcomp · Sep 16, 2004
| Started | Recall | Status |
|---|---|---|
| Jun 28, 2023 | Z-2377-2023, Z-2371-2023, Z-2378-2023, Z-2380-2023, Z-2390-2023, Z-2379-2023, Z-2376-2023, Z-2368-2023, Z-2372-2023, Z-2385-2023, Z-2388-2023, Z-2391-2023, Z-2367-2023, Z-2381-2023, Z-2383-2023, Z-2375-2023, Z-2374-2023, Z-2387-2023, Z-2382-2023, Z-2384-2023, Z-2369-2023, Z-2389-2023, Z-2370-2023, Z-2386-2023, Z-2373-2023Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully,… | Open |
- Power Acute Triple Lumen Hemodialysis CatheterK222170 · Health Line International Corporation · Jan 13, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 10, 2002
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.5540
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files