M3290B Intellivue Information Center Software, Release A.0
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K102495, from Philips Medical Systems (Andover, MA). Product code MHX, class 2.
As of Nov 18, 2026, K102495 is an FDA 510(k) held by Philips Medical Systems, substantially equivalent on Sep 30, 2010 after 30 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 30 days
- Recalls: 7
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFIndicated for central monitoring of multiple adult, pediatric, and neonatal patients; and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric, and neonatal patients and/or ST segment of adult patients to gain information for treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- M3290A Intellivue Information Center Software, Release 1.00K081983 · Philips Medical Systems · Aug 22, 2008
- M3290A Intellivue Information Center Software, Release J.00 (For M3140, M3145, M3150, M3151, M3154, M3155, M3169, M3170)K062271 · Philips Medical Systems · Nov 30, 2006
- M3290A Intellivue Information Center Software Release G.00K050742 · Philips Medical Systems · Apr 5, 2005
- M3290A Intellivue Information Center Software Release F.0 and M4840A Philips Telemetry System Ii Including M4841A PatienK041741 · Philips Medical Systems, Inc. · Jul 21, 2004
- M3290A Intellivue Information Center Software Release E.01 and Intellivue Clinical Network On VlanK040955 · Philips Medical Systems, Inc. · May 12, 2004
- M3290A Intellivue Information Center Software Release F.O and The M4840A Telemetry Monitoring System IiK040357 · Philips Medical Systems, Inc. · Mar 3, 2004
- M3290A Intellivue Information Center Software Release E.1K031403 · Philips Medical Systems, Inc. · May 30, 2003
- M3290A Intellivue Information Center Software Release E.OK023698 · Philips Medical Systems · Nov 19, 2002
- Philips Information Center Software Revision D.02K021422 · Philips Medical Systems · May 15, 2002
- Agilent Information Center Software for M3150,M3151,M3153 and M3154K011093 · Agilent Technologies, Inc. · May 1, 2001
- Agilent Information Center Software for M3150A and M3153A and Agilent M2/M3/M4 Compact Portable Patient MonitorK001057 · Agilent Technologies, Inc. · May 3, 2000
- Viridia Information Center Software for M3150A and M3153A, Viridia Telemetry System TransmitterK000854 · Agilent Technologies, Inc. · Apr 10, 2000
- Viridia Information Center Software for M3154A Opt C22K993907 · Agilent Technologies, Inc. · Dec 6, 1999
- HP Viridia Information System, Models M3150A/ M3151A/ M3153A/ M3154AK993171 · Hewlett-Packard Co. · Oct 21, 1999
- Hewlett-Packard Centralvue SoftwareK964832 · Hewlett-Packard Co. · May 15, 1997
Cited as a predicate
1 later 510(k) summary names this device as a predicate- M3290B Philips IntelliVue Information Center iXK143057 · Philips Medical Systems · Dec 3, 2014
Recalls
Recalls FDA ties to K102495, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Jan 3, 2018 | Z-0776-2018Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the… | Terminated |
| Nov 25, 2015 | Z-0858-2016, Z-0857-2016, Z-0856-2016Reconstructed ECG leads viewed or printed at the Information Center iX may misrepresent the ECG waveform in specific leads. | Terminated |
| Dec 19, 2014 | Z-0374-2016Sp02 and/or Non Invasive Blood Pressure (NBP) alarms may become disabled without visual notification | Terminated |
| Mar 5, 2013 | Z-1783-2014If a customer creates customized trend scales in the trend review tile and the iX or primary server reboots for any reason, the iX(s) will enter a reboot loop. If one iX reboots, the issue will impact only those… | Terminated |
| Nov 9, 2012 | Z-0495-2013Internal audio cable may have diminished retention force due to damaged contacts, which could lead to intermittent or loss of audible alarms. | Terminated |
Other MHX clearances
The newest 510(k)s for the same product code; their median review was 207 days; see every MHX device- 12SL PrecisionK261307 · Ge Medical Systems Information Technologies, Inc. · Nov 3, 2026
- Radius VSM and AccessoriesK260676 · Masimo Corporation · Oct 14, 2026
- Carescape ONE; Carescape Dock F0; TruSignal ECG MM; TruSignal PRES MM; TruSignal TEMP MM; TruSignal SpO2 MM 1.8M;…K260656 · GE Healthcare Finland Oy · Sep 1, 2026
- Carescape Canvas 1000; Carescape Canvas Smart DisplayK260343 · GE Healthcare Finland Oy · Aug 28, 2026
- Infinity® Acute Care System (IACS) – VG8.1K260579 · Draeger Medical Systems, Inc. · Aug 20, 2026
- Telemetry Monitor (iT30, iT50)K251508 · Edan Instruments, Inc. · Aug 6, 2026
- Monitor B105M; Monitor B125M; Monitor B155M; Monitor B105P; Monitor B125PK253277 · Ge Medical Systems Information Technologies, Inc. · Jul 24, 2026
- Radius VSM and AccessoriesK254209 · Masimo Corporation · Jun 4, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Aug 31, 2010
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1025
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files