M3290A Intellivue Information Center Software, Release J.00 (For M3140, M3145, M3150, M3151, M3154, M3155, M3169, M3170)

Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K062271, from Philips Medical Systems (Andover, MA). Product code MHX, class 2.

As of Nov 18, 2026, K062271 is an FDA 510(k) held by Philips Medical Systems, substantially equivalent on Nov 30, 2006 after 115 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 115 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Indicated for central monitoring of multiple adult, pediatric, and neonatal patients; and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric, and neonatal patients and/or ST segment of adult patients to gain information for treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K062271, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 17, 2010Z-2528-2017Potential for early component failure.Terminated
Nov 13, 2007Z-0451-2008Alarm failure : If changes are made to the Caregroup Editor configuration, Care Group alarm status, including alarm text and prompt tones, may fail to be displayed or annunciated at bedside monitors within the Care GroupTerminated
Jul 16, 2007Z-1182-2007Incorrect Patient Assignment: the information center will in most cases be associated with the "incorrect" patient. This occurs on both standalone and networked systems running fixed mode monitoring. (This does not…Terminated

Other MHX clearances

The newest 510(k)s for the same product code; their median review was 207 days; see every MHX device

Record

Type
Special
Received
Aug 7, 2006
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files