System 1E Liquid Chemical Sterilant Processing System

Sterilant, Medical Devices. 510(k) K102462, from STERIS Corporation (Mentor, OH). Product code MED, class 2.

As of Nov 26, 2026, K102462 is an FDA 510(k) held by STERIS Corporation, substantially equivalent on Sep 21, 2010 after 25 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 25 days
  • Recalls: 5
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K102462, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 21, 2013Z-0616-2014Customers indicated fluctuations in environmental conditions sites and gradual degradation/deformation of check valve o-rings following frequent contact with S40 sterilant use solution result in nuisance aborts.Terminated
Aug 1, 2012Z-2335-2012The firm initiated a recall after complaint investigations revealed that when the units are left on and not in use at night and/or on weekends the hose disconnects/separates resulting in water leakage that can cause…Terminated
Dec 7, 2011Z-0869-2012Software deficiency. STERIS identified that water specifications were not being met at some of their customer sites. Critical parameters for initiating cycles are: water quality, water pressure and water temperature.…Terminated
Aug 11, 2011Z-3124-2011STERIS learned that SYSTEM 1E users may experience aborted cycles due to a failure within the UV light system.Terminated
Mar 14, 2011Z-1960-2011The printout, display and Operator Manual for the STERIS System 1E Liquid Chemical Sterilant Processing System are subject to recall/correction by the firm because in the event of an incomplete or aborted cycle, the…Terminated

Other MED clearances

The newest 510(k)s for the same product code; their median review was 62 days; see every MED device

Record

Type
Special
Received
Aug 27, 2010
Review panel
General Hospital
Regulation
21 CFR 880.6885

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files