System 1.5 Sterile Processing System
Sterilant, Medical Devices. 510(k) K090036, from STERIS Corporation (Mentor, OH). Product code MED, class 2.
As of Nov 25, 2026, K090036 is an FDA 510(k) held by STERIS Corporation, substantially equivalent on Apr 5, 2010 after 454 days, with 6 recalls tied to it in FDA's recall database.
- Review time: 454 days
- Recalls: 6
- Cited as a predicate by: 5 devices
The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities.
- Steris System 1 Processor and 20 SterilantK875280 · Steris C/O Medical Device Consultants, Inc. · Sep 13, 1988
- Sterilox Liquid Chemical Sterilant SystemK013280 · Sterilox Technologies, Inc. · Sep 18, 2002
- System 1 endo Liquid Chemical Sterilant Processing SystemK173256 · STERIS Corporation · Mar 2, 2018
- System 1E Liquid Chemical Sterilant Processing SystemK131078 · STERIS Corporation · Jul 5, 2013
- System 1E Liquid Chemical Sterilant Processing SystemK113520 · STERIS Corporation · Feb 21, 2012
- System 1E Liquid Chemical Sterilant Processing SystemK102462 · STERIS Corporation · Sep 21, 2010
- System Ie Liquid Chemical Cterilant Processing SystemK101409 · STERIS Corporation · Aug 2, 2010
| Nov 21, 2013 | Z-0616-2014Customers indicated fluctuations in environmental conditions sites and gradual degradation/deformation of check valve o-rings following frequent contact with S40 sterilant use solution result in nuisance aborts. | Terminated |
| Aug 10, 2012 | Z-2284-2012STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick… | Terminated |
| Aug 1, 2012 | Z-2335-2012The firm initiated a recall after complaint investigations revealed that when the units are left on and not in use at night and/or on weekends the hose disconnects/separates resulting in water leakage that can cause… | Terminated |
| Dec 7, 2011 | Z-0869-2012Software deficiency. STERIS identified that water specifications were not being met at some of their customer sites. Critical parameters for initiating cycles are: water quality, water pressure and water temperature.… | Terminated |
| Aug 11, 2011 | Z-3124-2011STERIS learned that SYSTEM 1E users may experience aborted cycles due to a failure within the UV light system. | Terminated |
| Mar 14, 2011 | Z-1960-2011The printout, display and Operator Manual for the STERIS System 1E Liquid Chemical Sterilant Processing System are subject to recall/correction by the firm because in the event of an incomplete or aborted cycle, the… | Terminated |
- Endox High Level DisinfectantK254094 · Whiteley USA, Inc. · Oct 16, 2026
- SYSTEM 1E Liquid Chemical Sterilant Processing System, Model P6700; SYSTEM 1 endo Liquid Chemical Sterilant Processing…K251666 · STERIS Corporation · Apr 8, 2026
- SYSTEM 1E Liquid Chemical Sterilant Processing System (P6500); SYSTEM 1 endo Liquid Chemical Sterilant Processing…K232914 · STERIS Corporation · Oct 3, 2023
- SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900K230582 · STERIS Corporation · Mar 28, 2023
- SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model…K222615 · STERIS Corporation · Sep 28, 2022
- SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model…K211607 · STERIS Corporation · Sep 7, 2021
- SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model…K210737 · STERIS Corporation · May 12, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 6, 2009
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.6885
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files