Xio Rtp System
System, Planning, Radiation Therapy Treatment. 510(k) K102216, from Computerized Medical Systems, Inc. (Maryland Heights, MO). Product code MUJ, class 2.
As of Nov 19, 2026, K102216 is an FDA 510(k) held by Computerized Medical Systems, Inc., substantially equivalent on Oct 1, 2010 after 56 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 56 days
- Recalls: 7
- Cited as a predicate by: 2 devices
- Xio Rtp SystemK092132 · Computerized Medical Systems, Inc. · Sep 24, 2009
- Modification To Eclipse Treatment Planning SystemK091492 · Varian Medical Systems, Inc. · Jun 18, 2009
- Ray StationK140187 · Raysearch Laboratories AB · May 15, 2014
- Pinnacle3 Radiation Therapy Planning SystemK130992 · Philips Medical Systems (Cleveland), Inc. · Jun 14, 2013
| Aug 31, 2015 | Z-0130-2016Incorrect Treatment Delivery Using Third Party Fixed Wedges. | Terminated |
| Feb 17, 2014 | Z-1197-2014Dose and Monitor Unit values are not computed correctly when Elekta Motorized Wedges are used with Dynamic Conformal arcs. | Terminated |
| Nov 25, 2013 | Z-0780-2014When the number of fractions is changed for a proton spot beam, the MU are recomputed and rounded if necessary, but the dose is not recalculated as it should be even after saving and re-opening the plan. | Terminated |
| Nov 25, 2013 | Z-0928-2014When bolus is present, the Effective Depth (with bolus) to the weight point returned on the Source Data Report is incorrect. | Terminated |
| Apr 30, 2012 | Z-1644-2012XiO will underestimate the dose delivered when the closed leaf gap is set to 0mm on XiO and a value other than 0mm is set on the accelerator. | Terminated |
| Jun 9, 2011 | Z-2919-2011The XiO Software currently prompts for entry of CT to ED data for a studyset. No indication is given that the Electron Density data should only be used for Photon and Electron calculations and either Relative Stopping… | Terminated |
| Nov 16, 2010 | Z-2582-2011XiO: When a new electron beam is added to a plan, the calculation Algorithm is set to Pencil Beam, and an Aperture is added to the beam, the user can select Port-Port Properties and edit the Material Thickness values.… | Terminated |
- RT Elements (4.6)K261655 · Brainlab SE · Oct 5, 2026
- Virtual Cone (1.0)K260750 · Varian Medical Systems, Inc. · Jul 8, 2026
- RadOncAIK253050 · Informai · Jun 18, 2026
- Multimodality Simulation Workspace (MM Sim) (v1.0.0)K260207 · Philips Medical Systems Nederland B.V. · May 27, 2026
- eXaSkin PlusK252784 · Anatomical Geometry S.L. · May 22, 2026
- ART.1-USK253078 · Seetreat Pty, Ltd. · May 14, 2026
- syngo.via RT Image Suite VC10K252304 · Varian Medical Systems, Inc. · Mar 18, 2026
- Monaco RTP System (6.3)K252002 · Elekta Solutions AB · Feb 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 6, 2010
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files