Arca
Skip to content
Tools
Tools

Xio Rtp System

System, Planning, Radiation Therapy Treatment. 510(k) K102216, from Computerized Medical Systems, Inc. (Maryland Heights, MO). Product code MUJ, class 2.

As of Nov 19, 2026, K102216 is an FDA 510(k) held by Computerized Medical Systems, Inc., substantially equivalent on Oct 1, 2010 after 56 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 56 days
  • Recalls: 7
  • Cited as a predicate by: 2 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K102216, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 31, 2015Z-0130-2016Incorrect Treatment Delivery Using Third Party Fixed Wedges.Terminated
Feb 17, 2014Z-1197-2014Dose and Monitor Unit values are not computed correctly when Elekta Motorized Wedges are used with Dynamic Conformal arcs.Terminated
Nov 25, 2013Z-0780-2014When the number of fractions is changed for a proton spot beam, the MU are recomputed and rounded if necessary, but the dose is not recalculated as it should be even after saving and re-opening the plan.Terminated
Nov 25, 2013Z-0928-2014When bolus is present, the Effective Depth (with bolus) to the weight point returned on the Source Data Report is incorrect.Terminated
Apr 30, 2012Z-1644-2012XiO will underestimate the dose delivered when the closed leaf gap is set to 0mm on XiO and a value other than 0mm is set on the accelerator.Terminated
Jun 9, 2011Z-2919-2011The XiO Software currently prompts for entry of CT to ED data for a studyset. No indication is given that the Electron Density data should only be used for Photon and Electron calculations and either Relative Stopping…Terminated
Nov 16, 2010Z-2582-2011XiO: When a new electron beam is added to a plan, the calculation Algorithm is set to Pencil Beam, and an Aperture is added to the beam, the user can select Port-Port Properties and edit the Material Thickness values.…Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Record

Type
Traditional
Received
Aug 6, 2010
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files