Xio Rtp System

System, Planning, Radiation Therapy Treatment. 510(k) K092132, from Computerized Medical Systems, Inc. (Maryland Heights, MO). Product code MUJ, class 2.

As of Nov 19, 2026, K092132 is an FDA 510(k) held by Computerized Medical Systems, Inc., substantially equivalent on Sep 24, 2009 after 71 days, with 12 recalls tied to it in FDA's recall database.

  • Review time: 71 days
  • Recalls: 12
  • Cited as a predicate by: 2 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K092132, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 3, 2010Z-2579-2011XiO: When the beam template is recalled, XiO prematurely initiates a dose calculation, re-snaps the MLC or Port to the structure of the current patient , but then fails to re-calculate dose for the new MLC or Port…Terminated
Sep 10, 2010Z-2408-2011When RT Plans are exported from XiO and RT Structure Sets and Images are exported from Monaco, the DICOM iso-center coordinates may not be accurate.Terminated
May 28, 2010Z-2595-2011XiO Software: when Digitally Reconstructed Radiographs (DRRs) are generated in XiO, and DICOM is used to export the DRRs (File-Print-DRR), the exported images are associated with the Beams, but NOT the Plan. The Plan…Terminated
May 27, 2010Z-2593-2011XiO Software: The Patient Scan position is being confused with the treatment orientation of the patient. The Patient Scan position is determined from the data entered in PFM. Subsequently, when a New Teletherapy Plan is…Terminated
May 27, 2010Z-2407-2011XiO Software: For a specific clinical setup, the MLC leaf positions on the exported DRR images might not match the MLC leaf positions displayed in XiO.Terminated
May 19, 2010Z-2604-2011XiO Software: when a treatment machine is defined in Source File Maintenance (SFM), the MLC Model is set to Elekta, and beams using this treatment machine are added to a new plan in Teletherapy, XiO automatically adds…Terminated
Apr 15, 2010Z-1655-2010Under certain circumstances, XiO will over-estimate dose in the region where the backup jaw is inside the MLC openingTerminated
Feb 22, 2010Z-1209-2010Inconsistent MLC positions, dose, and monitor unitsTerminated
May 13, 2009Z-2568-2011In XiO, when the beams are renumbered, it is possible to get into a state where the dose displayed for a beam does not match the beam placement on the patient. This can occur when dose is calculated, beams are…Terminated
Dec 9, 2008Z-2567-2011XiO Software: Under certain conditions, when the port shape is altered, dose is not recalculated. The port-change flag is not being set correctly when the port shape is regenerated. The dose distribution will not match…Terminated
May 14, 2008Z-2574-2011In XiO, when Telepathy is exited, re-entered, and the temporary plan re-called, the OLD beam weight will be displayed but the dose distribution will reflect the modified beam weight.Terminated
Apr 8, 2008Z-2577-2011XiO: When the patient is positioned with their feet toward the CT gantry, the coordinates exported to the LAP IsoMark laser positioning system are not correct.Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Record

Type
Traditional
Received
Jul 15, 2009
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files