Zyno Z-800 Infusion System Model Z-800
Pump, Infusion. 510(k) K100705, from Zyno Medical, LLC (Waltham, MA). Product code FRN, class 2.
As of Nov 20, 2026, K100705 is an FDA 510(k) held by Zyno Medical, LLC, substantially equivalent on Jul 8, 2010 after 118 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 118 days
- Recalls: 8
- Cited as a predicate by: 2 devices
The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood and blood products to a human patient under the direction or supervision of physician or other certified health care professional.
- Z-800 Volumetric Infusion PumpK071545 · Zyno Medical, LLC · Oct 31, 2007
- Infusomat PK003029 · B.Braun Medical, Inc. · Dec 4, 2000
- Sigma 8000 and Sigma 8002K950766 · Sigma Intl. General Medical Apparatus, Inc. · Mar 22, 1995
- Acclaim Infusion PumpK991501 · Abbott Laboratories · May 13, 1999
- Z-800 Infusion SystemK130690 · Zyno Medical, LLC · Jun 12, 2013
- Zyno Medical Administration SetK120685 · Zyno Medical, LLC · Sep 7, 2012
| May 7, 2025 | Z-1868-2025, Z-1867-2025, Z-1866-2025, Z-1865-2025Unreleased software versions were installed on distributed devices without verification or validation. | Open |
| Oct 18, 2024 | Z-0506-2025Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains. | Open |
| Sep 30, 2024 | Z-0523-2025For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024. | Open |
| Sep 13, 2024 | Z-0005-2025There is a defect in the air-in-line software algorithm. | Open |
| Jun 13, 2011 | Z-1024-2013Update Instruction For Use for Preventative Maintenance (PM) Schedule and addition of Odometer software to monitor accumulative fluid volume infused by the pump | Terminated |
- RemunityPRO+ Pump for Remodulin® (treprostinil) InjectionK253703 · Deka R&D Corp · Jul 24, 2026
- FreedomEdge Infusion System; High-Flo SubQ Needle Set; Precision Flow Rate TubingK252015 · Koru Medical Systems, Inc. · Jan 28, 2026
- Spectrum IQ Infusion System with Dose IQ Safety Software (3570009)K251636 · Baxter Healthcare Corporation · Jul 28, 2025
- SIGMA Spectrum Infusion Pump with Master Drug LibraryK251640 · Baxter Healthcare Corporation · Jul 28, 2025
- MRidium 3870 MRI Infusion Pump System (3870)K242752 · Iradimed Corporation · May 28, 2025
- BD Alaris Infusion System with Guardrails Suite MXK243855 · Carefusion 303, Inc. · Apr 25, 2025
- Plum Solo™ Precision IV PumpK242114 · Icu Medical, Inc. · Apr 2, 2025
- Plum Duo™ Precision IV PumpK242115 · Icu Medical, Inc. · Apr 2, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Abbreviated
- Received
- Mar 12, 2010
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5725
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files