Arca
Skip to content
Tools
Tools

Xvi R4.5

Accelerator, Linear, Medical. 510(k) K100115, from Elekta , Ltd. (Crawley W. Sussex, GB). Product code IYE, class 2.

As of Nov 20, 2026, K100115 is an FDA 510(k) held by Elekta , Ltd., substantially equivalent on Mar 10, 2010 after 54 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 54 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Elekta X-Ray Volume Imaging system, XVI, is an electronic imaging device (EID), designed to be used with the Elekta range of medical linear accelerators and intended to be used as part of radiation therapy treatment process, as determined by a licensed medical practitioner.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • Xvi R5.0K131965 · Elekta Limited · Sep 25, 2013

Recalls

Recalls FDA ties to K100115, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 25, 2013Z-1059-2014XVI can incorrectly calculate the target position of the treatment table.Terminated
Jan 26, 2012Z-1066-2012If a user, using an on-line correction protocol, clicks the accept correction button, an approval dialog box appears. If cancel is selected, the table correction values return to their initial values, but the image…Terminated
Oct 20, 2011Z-0342-2012A small number of sites have experienced situations with kV generator transformers overheating due to a diode failure on the interface board.Terminated
Oct 14, 2011Z-0320-2012The software did not operate as expected.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Jan 15, 2010
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files