Elekta Synergy and 1 more
Accelerator, Linear, Medical. 510(k) K051932, from Elekta, Inc. (Norcross, GA). Product code IYE, class 2.
As of Nov 20, 2026, K051932 is an FDA 510(k) held by Elekta, Inc., substantially equivalent on Aug 12, 2005 after 30 days, with 14 recalls tied to it in FDA's recall database.
- Review time: 30 days
- Recalls: 14
- Cited as a predicate by: 3 devices
- Elekta Synergy SystemK032996 · Elekta Oncology Systems, Ltd. · Oct 23, 2003
- Precise Treatment System; Synergy platform; Synergy; Infinity; VersaHDK182138 · Eleckta Limited · Sep 5, 2018
- Xvi R4.5K100115 · Elekta , Ltd. · Mar 10, 2010
- Desktop ProK080585 · Elekta , Ltd. · Jun 11, 2008
| Dec 1, 2017 | Z-2840-2018There is a potential for an uncontrolled extension of iViewGT / XVI detector arms. | Terminated |
| Jul 8, 2016 | Z-2721-2016There have been instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts. | Terminated |
| Jan 7, 2015 | Z-1088-2015Elekta has identified a batch of securing bolts that fix the Gantry counter frame weight stack may fail. | Terminated |
| Nov 25, 2013 | Z-1061-2014Potential collision risk when using XVI and external beam shaping devices. | Terminated |
| May 28, 2013 | Z-1065-2014Customers are resetting the default values outside recommended factory settings. | Terminated |
| Dec 19, 2012 | Z-0931-2013The DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name. | Terminated |
| Aug 6, 2012 | Z-2405-2012There have been a small number of instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts. | Terminated |
| May 6, 2011 | Z-3079-2011A small number of the Modular Reverse Diodes (D1A), a component in the equipment room of the Elekta Linear Accelerator, have ruptured. | Terminated |
| Mar 9, 2011 | Z-0305-2012If "Confirm" or "Unconfirm" settings are clicked during the "Terminated Checking" phase of treatment delivery, Desktop Pro cancels the loaded beam and sets the Beam MUs to zero. Desktop Pro then sends the delivered MUs… | Terminated |
| Jan 5, 2011 | Z-1252-2011Flexmap calibration did not cover the full 360 degrees of gantry rotation (0 to 355 degrees) and that this caused an isocenter displacement with 2D PlanarView images required at 180 to -180. | Terminated |
| Oct 25, 2010 | Z-1723-2011If another patient is selected on the imaging system (XVI) release 4.5) during the transmission of end of treatment data from Desktop Pro R7.01 to R&V system, the XVI information is prioritized and the end of treatment… | Terminated |
| Sep 1, 2010 | Z-0350-2011It is possible to do couch corrections twice if both Couch Move Assistant in MOSAIQ and Remote Automatic Table in XVI are configured when in SYNERGISTIQ mode. | Terminated |
| Dec 12, 2008 | Z-1405-2010It is possible that two (or more) CT reference data sets are necessary for one patient, for example, if you have to scan a patient during breath hold and then again during free breathing. | Terminated |
| Sep 19, 2007 | Z-1411-2010Firm became aware of a failure in the position detection system which could result in incorrect positioning of the detector in the longitudinal or lateral direction. | Terminated |
- AerFrame™ Alignment Laser Box (LBV2)K261865 · Stabilix, LLC · Sep 16, 2026
- EZFluence (RADEZ V2)K261718 · Radformation, Inc. · Aug 31, 2026
- MRIA Radiation Therapy Information System (MRIA)K260733 · Manteia Technologies Co., Ltd. · Aug 26, 2026
- Halcyon, Ethos Radiotherapy System (5.0)K261856 · Varian Medical Systems, Inc. · Aug 3, 2026
- ThinkQA 3K252937 · Dosisoft SA · Jun 12, 2026
- Luna 3DK253004 · Lap GmbH Laser Applikationen · Jun 4, 2026
- AerFrame Patient Immobilization System (AFPIS)K260792 · Stabilix, LLC · May 8, 2026
- Veriqa RT Epid 3DK252258 · Ptw-Freiburg Physikalisch-Technische-Werkstaetten Dr. Pychla · Apr 15, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jul 13, 2005
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files